Instructions for use INTRA LUX Motor KL 703 LED – 1.007.0150
7 Reprocessing steps in accordance with ISO 17664-2 | 7.6 Sterilisation
23 / 28
7.6 Sterilisation
Sterilisation in a steam steriliser (autoclave) in
accordance with EN 13060 / EN ISO 17665-1
CAUTION
Contact corrosion due to moisture.
Damage to product.
4
Immediately remove the product from the steam steriliser after the steril-
isation cycle.
The medical device has a max. temperature resistance of up to 138
o
C (280.4
o
F).
Sterilisation parameters:
Select a suitable process from the following sterilisation processes (depending
on the available steriliser):
▪ Steriliser with triple pre-vacuum:
– at least 3 minutes at 134
o
C -1
o
C / +4
o
C (273
o
F -1.6
o
F / +7.4
o
F)
▪ Autoclave using the gravity method:
– at least 10 minutes at 134
o
C -1
o
C / +4
o
C (273
o
F -1.6
o
F / +7.4
o
F)
4
Use according to the manufacturer's Instructions for Use.
Ensure dryness. This is achieved by autoclaving with a subsequent vacuum, or
by a drying phase of 10 min with the autoclave door open.
NOTE
Before using it again, let the motor cool to room temperature.
7.7 Storage
Reprocessed products must be stored appropriately protected from dust in a
dry, dark and cool space with a low germ level.
NOTE
Comply with the expiry date of the sterilised items.