Instructions for use INTRA LUX Motor KL 703 LED – 1.007.0150
7 Reprocessing steps in accordance with ISO 17664-2 | 7.2 Manual reprocessing
21 / 28
CAUTION
Use of the disinfection bath or chloride-containing disinfectants.
Malfunction and material damage.
4
Do not disinfect the device in the disinfection bath or with chloride-contain-
ing disinfectants.
Approved disinfectants:
▪ CaviWipes and CaviCide made by Metrex
The validations of the efficacy and compatibility of materials were done using
the CaviWipes cleaning and disinfection wipes made by Metrex.
1
4
Do not take sleeve
①
off the motor.
Manual external cleaning
4
Thoroughly wipe down the outer surfaces of the medical device with
a CaviWipes cleaning and disinfection wipe to remove all visible deposits.
4
Thoroughly wipe down the outer surface of the medical device with a
second CaviWipes cleaning and disinfection wipe to remove any remaining
invisible contaminations.
Manual external disinfection
4
Finally, thoroughly wipe down the outer surfaces of the medical device with
a third CaviWipes cleaning and disinfection wipe and allow the agent to act
as detailed by the manufacturer of the disinfectant (3 min).
4
Follow the instructions for use of the disinfectant.
NOTE
Cleaning and servicing of the site of coupling.
To maintain uninterrupted operation, it is recommended to clean the grooves
for the O-rings semi-annually.
4
For this purpose, take all 3 O-rings off the coupling peg and remove all ad-
hering particles and lime scale from the grooves using a wet, lint-free cotton
cloth.
NOTE
If the connection to the instrument is not leak-proof, replace all O-rings.
7.2.2 Manual internal cleaning
Not applicable.
7.2.3 Manual internal disinfection
Not applicable.
7.2.4 Manual drying
CAUTION
Moisture.
Corrosion and other product damage.
4
Store motors in a dry place.