Instructions for use INTRA LUX Motor KL 703 LED – 1.007.0150

2 Safety | 2.8 Qualification of personnel

9 / 28

2.8 Qualification of personnel

Application of the product by users without the appropriate medical training
could injure patients, users or third parties.

4

Make sure that the user has read and fully comprehends the instructions for
use.

4

The device may be used only if the user has completed the appropriate
medical training.

4

Make sure that the user has read and comprehends the national and re-
gional regulations.

2.9 Service and repair

Repairs, servicing and safety checks may only be performed by trained service
personnel. The following persons are authorised to do this:

▪ Service technicians of KaVo branches after the appropriate product training

▪ Service technicians of KaVo authorised dealers after the appropriate product

training

Observe all the following items during servicing work:

4

Have the service and testing tasks carried out according to the Medical
Device Operator Ordinance.

4

After servicing, interventions on and repairs of the device and before re-use,
have the service personnel perform safety checks on the device.

4

KaVo recommends specifying in-house service intervals where the medical
device is brought to a professional shop for cleaning, servicing and a func-
tion check. Define the service interval depending on the frequency of use.

As a result of the use of NON-KaVo original spare parts during the repair, parts
such as covers may become undone and injure the patient, user or other
people. This may result in aspiration, swallowing of parts and possibly even a
risk of suffocation.

4

Only use spare parts that comply with the specification for repair; original
KaVo spare parts comply with the specification.

NOTE

If a repair is done with NON-KaVo original spare parts, this may constitute a
product modification that leads to the loss of CE conformity. In the event of
damage, the responsibility is with the service company or the operator.

The introduction of a modified product in the market, in which the reasonable
suspicion exists to endanger the safety and health of patients or users, is pro-
hibited by Medical Device Law §4, Abs.1 No. 1 and therefore requires its own
conformity check.

2.10 Property damage

Contaminated and moist compressed air can cause malfunctions and premature
wear.

4

Make sure to supply dry, clean and uncontaminated compressed air in ac-
cordance with DIN EN ISO 7494-2.

Contaminated water can cause malfunctions and premature wear.

4

Make sure that the water quality generally complies with DIN EN ISO
7494-2, permissible pH 7.2 to 7.8.

Содержание 1.007.0150

Страница 1: ...Instructions for use INTRA LUX Motor KL 703 LED 1 007 0150...

Страница 2: ...Distribution KaVo Dental GmbH Bismarckring 39 88400 Biberach Germany Phone 49 7351 56 0 Fax 49 7351 56 1488 Manufacturer KaVo Dental GmbH Bismarckring 39 88400 Biberach Germany www kavo com...

Страница 3: ...motor to the supply hose 16 5 2 Regulating the spray water 16 5 3 Attaching an instrument to the motor 16 5 4 Removing an instrument from the motor 17 6 Checking for malfunctions and troubleshooting 1...

Страница 4: ...ice If you have any questions or complaints please contact the KaVo Technical Service 49 0 7351 56 1000 service einrichtungen kavo com or service treatmentunits kavo com Please refer to the serial num...

Страница 5: ...es Transportation and storage conditions temperature range Transportation and storage conditions air pressure Transportation and storage conditions Humidity Protect from moisture Keep dry Protect from...

Страница 6: ...or use INTRA LUX Motor KL 703 LED 1 007 0150 1 User instructions 6 28 CAUTION In cases which if not prevented can lead to minor or moderate injury CAUTION In cases which if not prevented can lead to m...

Страница 7: ...ore disposal 2 2 Explosion hazard Electrical sparks in the product can lead to explosion or fire 4 Do not use the product in explosion hazard areas 4 Do not operate the product in an oxygen enriched a...

Страница 8: ...of time 2 6 Electromagnetic fields Electromagnetic fields might interfere with the functions of implanted systems such as pacemakers 4 Ask patients if they have a cardiac pacemaker or other system im...

Страница 9: ...als where the medical device is brought to a professional shop for cleaning servicing and a func tion check Define the service interval depending on the frequency of use As a result of the use of NON...

Страница 10: ...fety 2 11 Disposal 10 28 2 11 Disposal This product is subject to the EC directive governing waste electrical and elec tronic devices and must be submitted to special disposal in Europe 4 For further...

Страница 11: ...n areas at risk of explosion Intended only for dental treatment by a dental professional the product must not be modified or used for any other purpose since this may be haz ardous A medical device ac...

Страница 12: ...011 2600 ELECTROmatic TMM TMC Mat no 0 696 0161 KL 2 Mat no 1 004 4317 KL 3 Mat no 1 006 1037 KL 4 Mat no 1 003 3490 ELECTROtorque TLC Mat no 1 001 8748 ELECTROtorque PLUS Motor voltage rating 22 V AC...

Страница 13: ...duct to which the motor is to be attached Check the suitability of the product for operation with the motor NOTE Designed for operation on KaVo units 4 ESTETICA E80 ESTETICA E70 ESTETICA E50 ESTETICA...

Страница 14: ...pressure 700 hPa to 1060 hPa 10 psi to 15 psi 3 3 Transportation and storage conditions CAUTION Startup after refrigerated storage Malfunction 4 Prior to startup strongly refrigerated products must be...

Страница 15: ...to 7 Reprocessing steps in accordance with ISO 17664 Page 20 Currently applicable packaging law Dispose of and recycle the packaging appropriately in accordance with current packaging law employing wa...

Страница 16: ...speed is appropriate for the preparation on hand 4 Use the requisite minimal amount of spray water 4 Rotate the regulating sleeve of the motor hose to adjust the water portion of the spray in a contin...

Страница 17: ...ions for use INTRA LUX Motor KL 703 LED 1 007 0150 5 Operation 5 4 Removing an instrument from the motor 17 28 5 4 Removing an instrument from the motor 4 Pull the instrument off the motor in axial di...

Страница 18: ...n 4 Missing O ring of the motor coupling needs to be replaced 4 If the intensity of the KaVo Mini LED lamp is low increase the cold light in tensity on the unit 4 If the KaVo Mini LED lamp glows red o...

Страница 19: ...k properly Case 1 KaVo Mini LED lamp is on Case 2 The KaVo Mini LED lamp is faint 4 Increase the cold light intensity on the unit until the desired illuminance is reached Case 3 The KaVo Mini LED lamp...

Страница 20: ...films may be formed in the lines This is associated with a risk of infection This results in the following requirements for the workflow at the practice 4 Rinse all withdrawal sites of the water condu...

Страница 21: ...cal device with a second CaviWipes cleaning and disinfection wipe to remove any remaining invisible contaminations Manual external disinfection 4 Finally thoroughly wipe down the outer surfaces of the...

Страница 22: ...rvice regularly with suitable agents CAUTION Excess care products leaking from the contra angle handpiece oil cleanser and disinfectant can penetrate into the motor and cause it to malfunction 4 Pleas...

Страница 23: ...the following sterilisation processes depending on the available steriliser Steriliser with triple pre vacuum at least 3 minutes at 134 o C 1 o C 4 o C 273 o F 1 6 o F 7 4 o F Autoclave using the gra...

Страница 24: ...or KL 703 LED 1 007 0150 8 Consumables 24 28 8 Consumables Available from specialised dental dealers Material summary Mat No KaVo Mini LED lamp 1 007 8474 O ring 8 3x0 68 0 200 6120 Airflow measuring...

Страница 25: ...or defects and consequences thereof that have arisen or may arise from natural wear improper handling cleaning servicing or main tenance non compliance with operating maintenance or connection instruc...

Страница 26: ......

Страница 27: ......

Страница 28: ...1 008 1048 bd 20220914 06 en...

Отзывы: