4-8 INITIAL OPERATION
_______________________________________________________________________
Copyright © 2010 by Kaufman & Robinson, Inc., 1306 Blue Spruce Drive, Unit A, Fort Collins, CO 80524
Tel: 970-495-0187, Fax: 970-484-9350, Internet: www.ionsources.com
dicated in the Emission Amps display. Note that the Emission cur-
rent in amps is usually set equal to or up to 10% greater than the
discharge current.
• Use the white Select button on the Discharge Power Supply to se-
lect Volts. Turn the knob on the Discharge Power Supply until the
discharge voltage is set to maximum as indicated on the Discharge
Volts display.
• Use the white Select button on the Discharge Power Supply to se-
lect Amps. Turn the knob on the Discharge Power Supply until the
discharge current is set to 2.5 amps as indicated on the Discharge
Amps display.
• Press the white Enable/Standby button on the Keeper Power Sup-
ply. And wait a couple of seconds for the keeper to start the hollow
cathode as indicted by the Keeper Amps display reading 1.5 amps.
• Reduce the
fl
ow to gas channel 4 to 10 sccm.
• Press the white Enable/Standby button on the Emission Power Sup-
ply.
• Press the white Enable/Standby button on the Discharge Power
Supply.
• Adjust the gas
fl
ow for channel 1 using the Gas Adjust Knob on the
Auto Controller until the discharge voltage is approximately 120 V.
• Operate the ion source for at least 10 minutes to clean any contami-
nates from the ion source that may have been introduced while at
atmosphere.
• The discharge voltage will vary slightly during this time. Adjust the
gas
fl
ow after the 10 minutes to obtain a discharge voltage of ap-
proximately 120 V.
• At this point the operating conditions can be saved as a program to
be used later in the Auto Gas mode or the Manual Gas mode. To do
this press and hold the white Acquire Setpoints button for two sec-
onds. This saves the new parameters to the active program.
• Stop operation by putting the discharge, emission and keeper pow-