05.13 / V1.0
4-3
UNIDRIVE
®
S III
4.2 Notes to the safety checks
4.2.1. Tools and auxiliary material required for the safety checks
Power cord
Handpieces
Dual-pedal
footswitch
Order-No.
20
0166 30
Safety tester for electrical safety
4.2.2. Visual inspection
Tensioning lever pump
D
Open and close the tensioning lever of the pump several times.
✓
Limit with clear end points
✓
Locking levers are secure on the axis, no sliding/slipping when pressed against the limit
✓
No unusual grinding noises
4.2.3. Test for proper operation TFT devices
D
Connect the power cord and switch the device on.
✓
While switched on the power switch has to be illuminated.
✓
The error message E11
No footswitch
is displayed.
D
Connect the footswitch.
✓
The error message E12
No motor
is displayed.
D
Connect a handpiece to the motor socket M1.