8. Troubleshooting
55
Symptom Possible
Cause
Maintenance
【
LD ERR
】
The laser may be damaged.
Return the instrument to your distributor or to your
KANOMAX service center for repair.
The flow channel may be interrupted.
Make sure that the inlet or outlet is not blocked.
【
FLOW ERR
】
The pump may be damaged.
Return the instrument to your distributor or to your
KANOMAX service center for repair.
【
BATTERY ERR
】
Power capacity is low.
If using a rechargeable battery, charge the battery.
If using AC line, the line may have problems.
【
OVER ERR
】
The instrument may be being used
outside the operating environment in the
specifications.
Inside the optical system may be dirty.
Attach the zero filter to the inlet and perform aging.
【
STORAGE
ERR
】
The remaining capacity of CF card is too
low.
Delete unnecessary data inside the CF card.
【
Alarm
】
Reading exceeds the Alarm setting value.
-
www.actoolsupply.com