3. Display Description and Operation Procedure
29
3.6.6 STANDARD
STANDARD Measurement Setting
Based on the STANDARD setting configured in
3.9.2 OPTION
, the setting items for CLASS and
items in EXTENTION tab varies.
In the BASIC tab;
Configure the setting of CLASS, SCHEDULE,
ROOM, LOCATION, CYCLES and SAMPLE T.
CLASS
Set the cleanliness level of the cleanroom to be measured. After configuring this
setting, the appropriate particle size for the configured CLASS can be selected.
SCHEDULE You can load a MAP file. If you select the file, the MAP function will be
activated and the MAP will be displayed.
ROOM
Input the size of the room to be measured.
LOCATION
Input the number of measurement point.
CYCLES
Input the measurement cycles.
SAMPLE T.
Input the sampling time.
The numeric value shown next to the each entry field is the minimum value based on CLASS and ROOM. Tap
[SET] to use the recommendation value.
In the EXTENTION tab;
Configure the ROOM STATUS, and COMMENT
settings.
[ROOM STATUS]
Select either
as-built
,
as-rest
or
operational
.
-
Select
as-built
when the cleanroom is setting up.
-
Select
as-rest
when the cleanroom function is off.
-
Select
operational
when the cleanroom function is on.
[COMMENT]
Tap the entry field to display a keypad. Enter
comment and tap [Enter].
Tap [APPLY] to activate the new settings in each tab.
Tap [OK] to activate all the new settings.
Tap [CANCEL] to discard the changes and close the
window.