23
OODMANUAL 0720
KAM CONTROLS, INC.
O P E R AT I O N C O N T I N U E D
FIG. 4-4
Output data OOD models serial numbers OOD-19-140 or lower
Column 1
– Message ID (for information only)
Column 2
– Time code (for information only)
Column 3
– Background signal: Shows the output value from the PD sensor when the light source
is off. Range 0 to 2,500
Column 4
– PD sensor signal: Shows the output value from the PD sensor when the light
source is on. Range background sensor signal value to 65,535
Column 5
– LED Off signal: Shows the value of the LED light source when it is off.
Range 0 to 150
Column 6
– LED On signal: Shows the value of the LED light source when it is on.
Range 500 to 1,023
Column 7
– For factory use only
Column 8
– PD sensor voltage: Range 28,000 to 45,000
Column 9
– Differential value between the PD sensor signal
and the background signal
Column 10
– Light source compensated PD sensor
signal
Column 11
– Offset level compensated PD
Sensor Signal
Column 12
– Oil output: Range 0 to
5,000 ppm (Max)
Column 13
– 4-20 mA Output Loop
Column 14
– Electronics tem-
perature: Range 0.0 to 70.0 °C
Column 15
– CRC Cyclic
Redundancy Check
(for information only)