LRWMANUAL 0918
KAM CONTROLS, INC.
22
CALIBRATION EXAMPLE
Manual samples water content:
1. 2.51%
2. 2.75%
3. 2.62%
Average = 2.63%
LRW screen shots water content:
1. 2.60%
2. 2.91%
3. 2.70%
Average = 2.74%
Difference of averages:
2.74% - 2.63% = 0.11%
According to the averages, the LRW is reading 0.11% high.
Current reading: 3.0%
To adjust the LRW, the reading needs to be lowered by 0.11%.
3.0% – 0.11% = 2.89%.
This is the new value to be entered into the LRW.
K A M L R W O P E R AT I O N c o n t i n u e d
ON-LINE CALIBRATION OF THE KAM
®
LRW
™
continued
Type the calculated percent of water (i.e. 1.58 for 1.58%) and click on the Send ASCII button. FIG. 4-10.
Calibration results will be displayed in the RealTerm menu. FIG. 4-11
10.
FIG. 4-10
FIG. 4-11
Calibration
results