Full digital mechanical sector scanning ultrasonic diagnostic instruments (Vet)                                                    User’s Manual V1.07 

- 16 - 

4. After the image is automatically frozen, or press

key on the keyboard (or press the freeze 

switch

on the probe) to freeze the image, the machine automatically displays four groups’ 

measurement data and the corresponding waveform; 

    5. Press

key, choose and view the #1, #2, #3, #4 corresponding to the test waveform and 

measurement data. 

Note:   

1. If AVG values are displayed in red, it indicates that the measurement error is larger, 

need to re-test. 

2. Only choose P2 site, you can measure the “Loin”. 
3. You can get accurate values of the three layers of backfat at the P2 site (the last rib). 

Many hogs have the thin third layer of backfat. As they increase in weight and age, the 
third layer becomes thicker and wider spread over the body. Measurement at  P1 site 
(the trailing edge of shoulder: about 4-5 rib) is more difficult and inconsistent. P3 site 
(the lumbosacral junction) will get accurate measurement for two layers of backfat but 
may not give an accurate measurement for the third layers due to the muscle tissue.       

7.3 Measurement of Swine’s lean percentage 

    1. Enter A mode freeze state; 

2. Put ultrasonic coupling gel evenly to the appropriate site of swine’s back (if necessary, cut off 

the hair at the measurement site and wet the skin with hot water, then apply light oil to wet the 
skin one minute or two), in order to make good contact between swine skin and sound window. 
The probe is placed against the skin of the swine and is kept perpendicular to the back, move 
the probe gently, make sure no air bubbles between the skin and probe. Keep the swine quiet 
during the test (you can feed the swine during the test); 

3.  Press

key  on  the  keyboard  (or  press  the  freeze  switch

on  the  probe)  to  unfreeze  the 

image. Press

or

key, choose the measurement site P2 (the tenth rib); 

    4. After the image is automatically frozen, or press

key on the keyboard (or press the freeze 

switch

on the probe) to freeze the image, to measure the backfat thickness and loin 

thickness (see the above the method of backfat measurement); 

    5. Then, press

or

key, adjust the “Weight” (range: 037~300kg), the lower right will 

automatically display the lean percentage. 

7.4 Measurement considerations 

    1. When measuring the older animal, if necessary, it needs to remove the hair at the site and wet 

the  skin  with  hot  water,  then  apply  light  oil  to  wet  the  skin  one  minute  or  two  before 
measuring.  Always  keep  the  probe  perpendicular  to  the  swine’s  back;  errors  may  result  if 
probe  is  at  an  angle.  If  chooses  the  inaccurate  site,  such  as  animal  shoulder  or  arm,  the 
measured result is not accurate. 

2. When the probe leaves the measuring site, it needs to reapply ultrasonic coupling gel evenly to 

continue measurement. 

3. Have a bad contact, first remove any possible manure, straw and excess hair, etc., which may 

cause a poor contact between the sound window and skin and leading to inaccurate measured 
results. 

4. Recommend using dedicated ultrasonic coupling gel, also can use vegetable oil. 

Содержание MSU3

Страница 1: ...KAIXIN MSU3 Full Digital Mechanical Sector Scanning Ultrasonic Diagnostic Instruments Vet User s Manual Xuzhou Kaixin Electronic Instrument Co Ltd...

Страница 2: ...esponsible for the safety reliability and performance in case of meeting all the following requirements 1 Assembly expansion readjustment improve and repair are all performed by professionals recogniz...

Страница 3: ...problems damages or losses due to re installation alteration or repair the system by non Kaixin designated personnel 6 This product is intended to provide clinical diagnostic data for the doctor The...

Страница 4: ...a potential danger or unexpected use condition that may result in light injury or property loss or affecting the use if not avoided 2 Safety Symbols Symbols Meaning Symbols Meaning Type B applied part...

Страница 5: ...lation and Check 5 5 1 System installation 6 5 2 Install Remove the battery 6 5 3 Connection to shutter release 7 5 4 Connection to power 7 5 5 Ultrasonic probe check before and after operation 7 5 6...

Страница 6: ...eeze 13 6 7 Image Data storage and viewing 13 6 7 1 Image data storage 13 6 7 1 1 Save image in B 2B mode 13 6 7 1 2 Save data in A mode 14 6 7 2 View image or data 14 Chapter 7 Fat thickness and Lean...

Страница 7: ...it approx 620g including battery and probe 1 2 Primary Functions 1 System preset function 2 Mode conversion function 3 Adjustment of near field far field total gain and dynamic range 4 Frequency conve...

Страница 8: ...ing ultrasonic diagnostic instruments Vet are composed of main unit and probe etc 2 2 Components name Figure MSU3 main unit sketch map 2 3 Function keys description SN Function keys Function 1 Image p...

Страница 9: ...Chapter 3 System Configuration 3 1 Standard Configuration 1 Main unit 1 unit 2 3 5MHz mechanical sector probe 1 PC 3 Power adapter 1 PC 4 Internal battery 1 PC 5 Power cable 1 PC 6 OTG U disk 1PC 7 V...

Страница 10: ...may result in explosion Warning Avoid using this equipment with high frequency electric knife high frequency therapy equipment or defibrillators and other electronic devices or may an electric shock...

Страница 11: ...nnect the grounding wire to any gas pipe or water pipe or it may cause bad grounding and danger of explosion 3 This equipment is not waterproof not use this equipment in place where liquid may into th...

Страница 12: ...ry Service personnel or operator can replace the battery Before replacing the battery please contact Kaixin Company replace the battery under the guidance of Kaixin Company Please use the battery prov...

Страница 13: ...heck before and after operation Before and after ultrasonic diagnosis to check if there are any exceptionally on the surface of the probe or cable jacket such as peeling cracks bulge or if the acousti...

Страница 14: ...possible damage or blockage 9 Clean the field and environment 5 6 2 Inspection after start up Check the following items after starting the machine 1 No abnormal voice strange smell and overheating app...

Страница 15: ...ystem Set menu 2 Press or key to move the cursor to Time item and then press or key to adjust the time 3 Press key to confirm the time setting and quit system setting interface 6 2 3 Probe protect tim...

Страница 16: ...ress key to confirm this setting and quit system setting interface Explanation The ear tag scanning function is to be upgraded 6 2 7 Menu time setting 1 Press key and then press key to enter System Se...

Страница 17: ...Repeatedly press key also to achieve the cycle conversion among the B A 2B mode 6 4 Image adjustment 6 4 1 Image adjustment in B 2B mode In B 2B mode and in real time status you can adjust images gain...

Страница 18: ...ress direction key to reduce far field gain so as to control the gain in far field region 3 After waitting for 5 30s Menu Time set in the System Set the parameters will be saved and exit the image adj...

Страница 19: ...e and in real time status press key to enter gain adjustment status 2 Press or key to adjust the gain within 27 90 3 After waitting for 5 30s Menu Time set in the System Set the parameters will be sav...

Страница 20: ...uary 14 2016 on the storage path U IMAGE save a 20160114 csv EXCEL spreadsheet 6 7 1 2 Save data in A mode 1 Insert OTG U disk into the Micro USB interface at the right side of main unit the lower lef...

Страница 21: ...propriate site of swine s back if necessary cut off the hair at the measurement site and wet the skin with hot water then apply light oil to wet the skin one minute or two in order to make good contac...

Страница 22: ...make good contact between swine skin and sound window The probe is placed against the skin of the swine and is kept perpendicular to the back move the probe gently make sure no air bubbles between the...

Страница 23: ...following table Glutaraldehyde based disinfectant Chemical Name Reagent Name Step Glutaraldehyde 2 4 Cidex Glutaraldehyde disinfectant Please refer to the instructions of the solution for details Warn...

Страница 24: ...ou may follow the recommended method by the disinfectant manufacturer to rinse 4 After finishing the rinsing use a sterilized cloth or gauze to wipe the water on the surface of probe Do not dry the pr...

Страница 25: ...g will shorten the battery life the full charged battery can be used about 5 hours 4 Battery is consumable the battery cycle life is based on the times of charge and discharge as unit When the use tim...

Страница 26: ...ch will lead to the leakage explosion or fire 12 Do not mixed use the battery together with one off battery such as dry battery or different capability or different model or different brand battery wh...

Страница 27: ...far field Control panel malfunction Turn off the machine and restart it 8 4 Periodic Safety Checks To ensure the system performance and safety it must be checked after using 1 year When check the inst...

Страница 28: ...nd corrosive gas 3 Shockproof measures have been taken inside the packing case to allow for transport by air railway land and sea The goods shall not be exposed to poor weather conditions like rain an...

Страница 29: ...zhou Jiangsu China Post Code 221004 Tel 86 516 87732932 87733758 Fax 86 516 87732932 87792848 Website http www kxele com E mail info kxele com Shanghai International Holding Corp GmbH Europe Eiffestra...

Отзывы: