Full digital mechanical sector scanning ultrasonic diagnostic instruments (Vet)                                                    User’s Manual V1.07 

- 14 - 

5. Press

key to return to the real-time status. 

Note 1: The “IMAGE” folder includes “images folder” and “EXCEL” spreadsheet. 
Note 2: The “images folder” is automatically named by the machine current date; the 

image is automatically named by the machine current time. 

Note 3: “EXCEL” spreadsheet is automatically named by the machine current date.   
Note 4: The EXCEL spreadsheet records the pregnancy status for each inspection 

according to chronological order. In the EXCEL spreadsheet, the contents of 
each column respectively: serial number, Ear tag, check date, check time, and 
pregnancy status.   

Note 5: For the ear tag function, to be upgraded. The Ear tag is expressed as 

A000000000000 in the “EXCEL” spreadsheet. 

                      Such as: 

U:\IMAGE\20160217\082736.bmp 

                      It represents that at 08:27:36

 

on Feburary 17, 2016 and on the storage path 

U:\IMAGE\20160217, save a “082736.bmp” image. 

 
                      U:\IMAGE\20160114.csv 
                      It represents that on January 14, 2016, on the storage path U:\IMAGE, save 

a “20160114.csv” EXCEL spreadsheet. 

6.7.1.2 Save data in A mode 

1. Insert OTG U disk into the Micro-USB interface at the right side of main unit, the lower left 

corner of the screen appears “U Disk” prompt; 

2. Complete the A-mode measurement, freeze the image; 

3. Press

key, the system issued a “tick” tone, namely the current data is stored in the 

EXCEL spreadsheet 

in the 

“DATA” folder of the U disk; 

4. Press

key to return to the real-time status. 

Note 1: The “DATA” folder includes “EXCEL” spreadsheet. EXCEL spreadsheet is 

automatically named by the machine current date. For example, the machine 
current time is January 18, 2016; the file name of EXCEL spreadsheet is 
20160118. 

Note 2: The EXCEL spreadsheet records the results for each measurement in the order. 

In the EXCEL spreadsheet, the contents of each column respectively: Ear tag, 
check date, check time, measurement site P1/P2/P3, the weight of swine, lean 
percentage, the backfat thickness

 

and loin thickness. 

Note 3: For the ear tag function, to be upgraded. The Ear tag is expressed as 

A000000000000 in the “EXCEL” spreadsheet. 

6.7.2 View image or data 

Insert OTG U disk to computer to view the stored images or data. 

 

 

 

 

 

 

 

 

Содержание MSU3

Страница 1: ...KAIXIN MSU3 Full Digital Mechanical Sector Scanning Ultrasonic Diagnostic Instruments Vet User s Manual Xuzhou Kaixin Electronic Instrument Co Ltd...

Страница 2: ...esponsible for the safety reliability and performance in case of meeting all the following requirements 1 Assembly expansion readjustment improve and repair are all performed by professionals recogniz...

Страница 3: ...problems damages or losses due to re installation alteration or repair the system by non Kaixin designated personnel 6 This product is intended to provide clinical diagnostic data for the doctor The...

Страница 4: ...a potential danger or unexpected use condition that may result in light injury or property loss or affecting the use if not avoided 2 Safety Symbols Symbols Meaning Symbols Meaning Type B applied part...

Страница 5: ...lation and Check 5 5 1 System installation 6 5 2 Install Remove the battery 6 5 3 Connection to shutter release 7 5 4 Connection to power 7 5 5 Ultrasonic probe check before and after operation 7 5 6...

Страница 6: ...eeze 13 6 7 Image Data storage and viewing 13 6 7 1 Image data storage 13 6 7 1 1 Save image in B 2B mode 13 6 7 1 2 Save data in A mode 14 6 7 2 View image or data 14 Chapter 7 Fat thickness and Lean...

Страница 7: ...it approx 620g including battery and probe 1 2 Primary Functions 1 System preset function 2 Mode conversion function 3 Adjustment of near field far field total gain and dynamic range 4 Frequency conve...

Страница 8: ...ing ultrasonic diagnostic instruments Vet are composed of main unit and probe etc 2 2 Components name Figure MSU3 main unit sketch map 2 3 Function keys description SN Function keys Function 1 Image p...

Страница 9: ...Chapter 3 System Configuration 3 1 Standard Configuration 1 Main unit 1 unit 2 3 5MHz mechanical sector probe 1 PC 3 Power adapter 1 PC 4 Internal battery 1 PC 5 Power cable 1 PC 6 OTG U disk 1PC 7 V...

Страница 10: ...may result in explosion Warning Avoid using this equipment with high frequency electric knife high frequency therapy equipment or defibrillators and other electronic devices or may an electric shock...

Страница 11: ...nnect the grounding wire to any gas pipe or water pipe or it may cause bad grounding and danger of explosion 3 This equipment is not waterproof not use this equipment in place where liquid may into th...

Страница 12: ...ry Service personnel or operator can replace the battery Before replacing the battery please contact Kaixin Company replace the battery under the guidance of Kaixin Company Please use the battery prov...

Страница 13: ...heck before and after operation Before and after ultrasonic diagnosis to check if there are any exceptionally on the surface of the probe or cable jacket such as peeling cracks bulge or if the acousti...

Страница 14: ...possible damage or blockage 9 Clean the field and environment 5 6 2 Inspection after start up Check the following items after starting the machine 1 No abnormal voice strange smell and overheating app...

Страница 15: ...ystem Set menu 2 Press or key to move the cursor to Time item and then press or key to adjust the time 3 Press key to confirm the time setting and quit system setting interface 6 2 3 Probe protect tim...

Страница 16: ...ress key to confirm this setting and quit system setting interface Explanation The ear tag scanning function is to be upgraded 6 2 7 Menu time setting 1 Press key and then press key to enter System Se...

Страница 17: ...Repeatedly press key also to achieve the cycle conversion among the B A 2B mode 6 4 Image adjustment 6 4 1 Image adjustment in B 2B mode In B 2B mode and in real time status you can adjust images gain...

Страница 18: ...ress direction key to reduce far field gain so as to control the gain in far field region 3 After waitting for 5 30s Menu Time set in the System Set the parameters will be saved and exit the image adj...

Страница 19: ...e and in real time status press key to enter gain adjustment status 2 Press or key to adjust the gain within 27 90 3 After waitting for 5 30s Menu Time set in the System Set the parameters will be sav...

Страница 20: ...uary 14 2016 on the storage path U IMAGE save a 20160114 csv EXCEL spreadsheet 6 7 1 2 Save data in A mode 1 Insert OTG U disk into the Micro USB interface at the right side of main unit the lower lef...

Страница 21: ...propriate site of swine s back if necessary cut off the hair at the measurement site and wet the skin with hot water then apply light oil to wet the skin one minute or two in order to make good contac...

Страница 22: ...make good contact between swine skin and sound window The probe is placed against the skin of the swine and is kept perpendicular to the back move the probe gently make sure no air bubbles between the...

Страница 23: ...following table Glutaraldehyde based disinfectant Chemical Name Reagent Name Step Glutaraldehyde 2 4 Cidex Glutaraldehyde disinfectant Please refer to the instructions of the solution for details Warn...

Страница 24: ...ou may follow the recommended method by the disinfectant manufacturer to rinse 4 After finishing the rinsing use a sterilized cloth or gauze to wipe the water on the surface of probe Do not dry the pr...

Страница 25: ...g will shorten the battery life the full charged battery can be used about 5 hours 4 Battery is consumable the battery cycle life is based on the times of charge and discharge as unit When the use tim...

Страница 26: ...ch will lead to the leakage explosion or fire 12 Do not mixed use the battery together with one off battery such as dry battery or different capability or different model or different brand battery wh...

Страница 27: ...far field Control panel malfunction Turn off the machine and restart it 8 4 Periodic Safety Checks To ensure the system performance and safety it must be checked after using 1 year When check the inst...

Страница 28: ...nd corrosive gas 3 Shockproof measures have been taken inside the packing case to allow for transport by air railway land and sea The goods shall not be exposed to poor weather conditions like rain an...

Страница 29: ...zhou Jiangsu China Post Code 221004 Tel 86 516 87732932 87733758 Fax 86 516 87732932 87792848 Website http www kxele com E mail info kxele com Shanghai International Holding Corp GmbH Europe Eiffestra...