Full digital mechanical sector scanning ultrasonic diagnostic instruments (Vet) User’s Manual V1.07
- 5 -
Chapter 5 System Installation and Check
Warning:
1.
When instrument works abnormally, do stop working, turn off the power and check
the reason, then contacts the KAIXIN Company about it.
2.
Turn off power and pull out of the plug from socket after each ultrasonic diagnostic
operation.
3.
It is forbidden to drag and press the power and probe cables emphatically; regularly
inspect whether there is pull-apart and bareness, if there is the phenomena like this,
turn off power supply immediately, stop using it and change it for new one.
4.
It is forbidden to load and unload the probe or move the instrument in galvanic to
avoid danger of safety.
5.
Pull out of the plug from socket after operation in thunderstorm weather to avoid
the instrument being damaged by lightening.
6.
If the temperature changes greatly in short time will cause vapor recovery inside of
instrument, the case may damage the instrument.
7.
The equipment is switched completely only by disconnecting the power supply from
the wall socket.
Warning:
To avoid the risk of electric shock, this equipment must only be connected
to a supply mains with protective earth.
Warning:
1.
All plugs of instruments of this system shall be connected into the power socket
with protectively earth on the wall and the socket must meet the requirement of
power rating of instrument. Use of multiple portable socket-outlets may affect
protective earth to make leakage currents exceed the safety requirements.
2.
Please follow the correct electrical connections method to connect the power supply
and earth, otherwise there will be danger of electric shock. Do not connect the
grounding wire to any gas pipe or water pipe, or it may cause bad grounding and
danger of explosion.
3. This equipment is not waterproof, not use this equipment in place where liquid
may into the interior of the equipment. Never pour any liquid on the equipment;
otherwise there will be danger of electric shock or cause equipment damage. If
accidentally spill liquid on the equipment, turn off the power immediately and
contact your local representative.
4. Additional equipment connected to the medical electrical equipment must comply
with the respective IEC or ISO standards (e.g. IEC60950 for data processing
equipment). Furthermore all configurations shall comply with the requirements
for medical electrical systems (see IEC 60601-1-1 or clause 16 of IEC60601-1 3rd,
respectively). Anybody connecting additional equipment to medical electrical
equipment configures a medical system and is therefore responsible that the
system complies with the requirements for medical electrical systems. Attention is
drawn to the fact that local laws take priority over the above mentioned
requirements. When more than one device is connected to use, the accumulation of
leakage current may cause the danger of
security. If in doubt, consult your local
representative or the technical service department.
5.
Prohibit the live parts of the equipment or other devices (such as various signal
input and output ports, etc.) contact with the patient, if this equipment or other
equipment has failure, the patient will have danger of electric shock.
6.
If the integrity of the external protective conductor in the installation or its
arrangement is in doubt, equipment shall be operated from its internal electrical
power source.
Warning:
The power adapter, shutter release, battery and power supply cord as
described in this section are replaced by operator. But these parts must be provided by
KAIXIN or his authorized supplier.