6 Electrical connection
64
COM 2
digital sensors
Connection to optional board:
RS422/485 serial interface
For connection of a digiLine
bus line, JUMO offers
5-wire M12 digiLine Master
connecting cable. Up to 6 sen-
sors with digiLine electronics
can be operated over a dig-
iLine bus. The voltage supplies
of DC 5 V and DC 24 V for the
sensors on the bus are provid-
ed by the voltage supply out-
puts on the device (base unit or
optional board).
⇨
Chapter 6.4.9 "Voltage sup-
ply outputs", page 59
Connecting the M12 digiLine
connecting cable:
A = Gray
B = Black
C = Brown
D = Blue
E = Blue (preassembled
jumper for connection of
the GND terminals for 5
and 24 V)
F = White
G = Shield (black with
cable lug for
connection to the ground)
with terminating resistors
without terminating resistors
Slot
Connection variant
Terminating resistors
Symbol
1
2
3
4
1
2
3
4
A
B
C
D
E
G
F
3
4
+5 V
GND
+24 V
GND