8 Operation
96
8.2.5
Service
The "Service" menu is used for in-device troubleshooting and diagnostics. It is
intended primarily for trained personnel. When troubleshooting together with
Service, the operator can access data that the JUMO service technicians need
for diagnosis.
Service Menu items
•
Default-configuration:
The operator can store the current device configu-
ration in the device's memory. This configuration can be reloaded as the ac-
tive configuration at any time. This is important, for instance, to return to the
initial configuration after configuration changes made for testing purposes.
Configurations can also be saved with the PC setup program or on a USB
flash drive with the aid of the Memory Manager.
⇨
see B 202581.0 in the "PC Setup Program" chapter
⇨
Chapter 8.4 "Memory Manager (USB flash drive)", page 108
•
Debug Window:
The Debug window can be used to check the behavior of
the device software in detail. Data are displayed to assist the JUMO service
technicians during troubleshooting.
•
Ci base calibration:
The Ci base calibration is required when starting up an
optional board or sensor for inductive conductivity measurements and can
be performed here.
⇨
Chapter 9.3 "Ci base calibration", page 121
•
Digital sensor 1 to 6:
For each digital JUMO sensor from the product group
2026xx that is operating on the device, a menu appears here that allows you
to reset the calibration date of the sensor to the default settings.
NOTE!
The Service menu appears in the Device menu only if a
user with corresponding user rights is logged in.
⇨
Chapter 8.2.1 "Logon/Logoff", page 91