113
9 Retrofitting optional boards
10
Install the M12 socket supplied with the optional board in a suitable
cable entry opening in the case.
11
Connect the 2-wire temperature sensor cable from the socket to an
appropriate analog input (e. g. temperature measurement input).
Observe the information regarding the integrated temperature sen-
sor in the conductivity sensor.
12
Insert the plug-in screw terminal for the M12 socket into the Ci op-
tional board. The factory-installed wiring must not be changed.
13
Connect the inductive conductivity sensor to the M12 socket.
14
Reinsert all other plug-in screw terminals and interface cables.
15
Reattach the terminal compartment cover.
Tighten all 6 screws in the terminal compartment cover to a torque
of 1 Nm.
16
Now, reconnect the power supply and check whether the new
hardware has been recognized.
⇨
Chapter 7.4.1 "Checking the optional boards", page 75
17
Only for Ci optional boards (inductive conductivity):
Perform a Ci base calibration.
⇨
Chapter "Performing the Ci base calibration", page 122
18
Only for optional boards for analysis inputs:
Calibrate the analysis inputs.
⇨
Chapter 11 "Calibration in general", page 45