6 Electrical connection
52
Electrolytic conductivity analysis inputs
Slot
Option/connection variant
Symbol
IN 7
IN 8
IN 9
IN 10
Ci optional board (inductive conductivity measurement)
Connection via an M12 plug,
Attach connections for compensation
thermometer (2-wire cable for socket) to a
suitable analog input (2-wire circuit),
The factory-installed wiring must not be changed!
Core colors of the conductor connection of the M12 socket to
the screw terminal connection on the optional board:
A = Brown
B = White
C = Pink
D = Silver
E = Black
F = Green (temperature sensor)
G = Yellow (temperature sensor)
CR optional board (resistive conductivity measurement)
2-electrode system with 2-wire conductor
Terminal 1 must be connected to the
outer electrode on concentric conductivity sensors.
A = Outer electrode (core color for JUMO types with fixed ca-
ble: White)
B = Inner electrode (core color for JUMO types with fixed ca-
ble: Brown)
C = Shield
1
2
3
4
6
Temp.
measur
input
A
G
B
C
D
E
F
C
B
A
1
2
3
4
6
Содержание AQUIS touch S 202581