background image

25

7 Installation

The choice of cable, the installation and the electrical connection must 
conform to the requirements of VDE 0100 “Regulations on the Installation 
of Power Circuits with Nominal Voltages below 1000 V” or the appropriate 
local regulations.

The electrical connection must only be carried out by qualified personnel.

If contact with live parts is possible while working on the device, it must be 
completely disconnected from the electrical supply.

The electromagnetic compatibility conforms to EN 61326.

Run input, output and supply cables separately and not parallel to one 
another.

The device is not suitable for use in areas with an explosion hazard (Ex 
areas).

Apart from faulty installation, incorrect settings on the instrument may also 
affect the proper functioning of the subsequent process or lead to damage. 

The electrical connection must only be carried out
by properly qualified personnel!

All manuals and user guides at all-guides.com

Содержание 202755

Страница 1: ...e Conductivity Concentration and Temperature Transmitter with switch contacts Type 202755 Operating Instructions 20275500T90Z001K000 V5 01 EN 00444870 All manuals and user guides at all guides com a l...

Страница 2: ...idiary or the head office in such a case Resetting the LC display If the brightness contrast setting is such that the text in the display is not reada ble the basic setting can be restored as follows...

Страница 3: ...ce version 9 4 1 Nameplate 9 4 2 Order details 10 5 Device description 13 5 1 Transmitter technical data 13 6 Mounting 18 6 1 General 18 6 2 Head mounted transmitter 19 6 3 Split version separate sens...

Страница 4: ...unction 47 11 Calibration 51 11 1 General 51 11 2 Calibrating the relative cell constant 51 11 3 Calibrating the temp coefficient of the sample solution 52 12 Maintenance 60 12 1 Cleaning the conducti...

Страница 5: ...tly Note This symbol is used when your special attention is drawn to a remark abc1 Footnote Footnotes are remarks that refer to specific points in the text Footnotes consist of two parts A marker in t...

Страница 6: ...suring transmitter Examples 1 Transmitter 4 Inductive conductivity sensor 2 Process connection 5 With or without graphics LC display 3 Exposed temperature sensor 2 1 3 Head mounted version Transmitter...

Страница 7: ...ugh the user friendly PC setup program Simply rotating the housing cover makes it possible to read the display regardless of whether the installation is in horizontally or vertically arranged pipes By...

Страница 8: ...put signals are electrically isolated from one another and from the medium of the measuring cell The conductivity is measured using an inductive probe A sinusoidal a c voltage feeds the transmitting c...

Страница 9: ...0465201018240001 The characters in question are in positions 12 to 15 from the left The device was therefore produced in the year 2018 in its 24th calendar week Description Description on the nameplat...

Страница 10: ...oard consisting of transmitter with permanently mounted sensor1 202755 15 JUMO CTI 500 Head transmitter with display keyboard 2 Process connection 168 PVC union nut G 1 A2 6 169 Stainless steel union...

Страница 11: ...07 Taper socket with union nut DN 50 DIN 11851 dairy compression fitting 617 Clamping socket clamp 2 similar to DIN 32676 3 690 SMS DN 2 706 immersion version 4 Immersion length 0 not available 0500 0...

Страница 12: ...ivery If required please order additionally 4 If required order extra code 580 5 Show type extension codes in succession separated by a comma 6 Only in case of process extension 168 and 169 in combina...

Страница 13: ...the solid state relays U DC 45 V U AC 30 V I 200 mA Electrical connection plug in screw terminals 2 5 mm2 or M12 plug socket connector Display option graphics LCD with background lighting adjustable...

Страница 14: ...A 20 4 mA The output signal is freely scalable Burden 500 for current output 2k for voltage output Analog output for Alarm Low 0 mA 0 V 3 4 mA 1 4 V or High 22 0 mA 10 7 V or a value with a fixed sett...

Страница 15: ...lue with a fixed setting safe value 5 1 4 Temperature compensation Reference temperature 15 to 30 C adjustable Temperature coefficient 0 0 to 5 5 C adjustable Compensation range 20 to 150 C Function L...

Страница 16: ...t the cell operating life Temperature of the sample medium Note the limiting values ambient temperature of the device Pressure 10 bar max at 20 C 06 bar max at 60 C 1 not compensated for temperature P...

Страница 17: ...s approval marks Approval mark DNV GL Testing agency DNV GL Certificate certification number TAA00001W9 Inspection basis DNV GL Class Guideline CG 0339 Valid for Type 202755 10 Type 202755 15 All manu...

Страница 18: ...of the pipe If it is not possible to achieve this minimum clearance then a limited compensation can be made through the Mounting factor parameter For submerged operation in basins a location must be...

Страница 19: ...19 6 Mounting 6 2 Head mounted transmitter Installation variations All manuals and user guides at all guides com...

Страница 20: ...6 Mounting 20 6 3 Split version separate sensor Transmitter head Drilling jig for wall mounting All manuals and user guides at all guides com...

Страница 21: ...21 6 Mounting Sensor component All manuals and user guides at all guides com a l l g u i d e s c o m...

Страница 22: ...on model Feststellschraube Flansch verschiebbar 60 18 3 100 5 0 18 125 165 Optional accessory Flange DN 32 part no 00083375 Optional accessory Flange DN 50 part no 00083376 Sliding flange Fixing screw...

Страница 23: ...No 20 00439249 Version with process connection 168 3URFHVV FRQQHFWLRQ 0 1 7KUHDGHG SLSH DGDSWRU 1 6 1 5HGXFHU 7HH WR EH SURYLGHG E WKH SODQW RSHUDWRU OPU TVQQMJFE CZ UIF EFWJDF NBOVGBDUVSFS 1 VKRUW 66...

Страница 24: ...ews 1 M5 x 30 are used for pipe diameters from 30 to 40 mm The screws 2 M5 x 40 are used for pipe diameters from 40 to 50 mm The pipe assembly set is also suitable for horizontal pipes 30 50 mm 2 1 1...

Страница 25: ...rking on the device it must be completely disconnected from the electrical supply The electromagnetic compatibility conforms to EN 61326 Run input output and supply cables separately and not parallel...

Страница 26: ...sary to open the housing for devices with cable glands Devices with M12 plug socket connectors should not be opened 1 To connect the single conductors pull off the pluggable screw terminals 1 in the o...

Страница 27: ...lease note that the serial number of the external sensor marked on the label attached to the connecting cable must match the serial number marked on the nameplate of the transmitter Caution On devices...

Страница 28: ...10 0 V or 2 10 V resp 10 2 V electrically isolated 3 4 I 3 I 4 Analog signal output Temperature 0 20 mA resp 20 0 mA or 4 20 mA resp 20 4 mA or 0 10 V resp 10 0 V or 2 10 V resp 10 2 V electrically i...

Страница 29: ...ons e g the dilution function Setting up a customer specific characteristic etc Connections Data transmission from or to the transmitter can only take place when it is connected to the electrical supp...

Страница 30: ...delivered ready for operation For devices with a separate sensor type code extensions 2 60 or 65 the transmitter and detached sensor are matched to one another at the factory When connecting the comp...

Страница 31: ...ue step on in selection 5 key reduce value step on in selection 6 LEDs K1 and K2 show the states of the switched outputs In normal operation the LED lights up if the corresponding output is active If...

Страница 32: ...has run down Calib customer calibration is active 7 8 9 10 11 Output mode Hand manual operation Hold hold operation Conductivity concentration measurement Unit for conductivity concentration measureme...

Страница 33: ...Y VALUE MANUAL OPERATION MANUAL VALUE INPUT TEMPERATURE UNIT MEAS VALUE ACQUISITION MANUEL SPECIFICATION OFFSET FILTER TIME OUTPUT TEMPERATURE SIGNAL TYPE SCALING START SCALING END IN CASE OF ALARM DU...

Страница 34: ...apter 10 6 2 CONDUCTIVITY OUT conducti vity output page 38 INPUT CONDUCTIVITY OUTPUT CONDUCTIVITY INPUT TEMPERATURE OUTPUT TEMPERATURE OUTPUT BINARY 1 OUTPUT BINARY 2 INPUT BINARY 1 INPUT BINARY 2 DIL...

Страница 35: ...35 10 Operation 10 4 Principle of operation Operation in levels All manuals and user guides at all guides com...

Страница 36: ...e been enabled by the administrator administrator level can be edited in this level All other parameters marked by a key can only be read Press the key for at least 3 seconds Select OPERATOR LEVEL 1 M...

Страница 37: ...EAR NON LINEAR see Non linear temperature coefficient ALPHA Page 55 NATURAL WATER permissible temperature range 0 to 36 C to EN 27 888 TEMPCO 1 4 1 0 2 20 5 5 1 Ranges 2 3 and 4 are only used if BINAR...

Страница 38: ...10 6 2 CONDUCTIVITY OUT conductivity output SIGNAL TYPE 0 20 mA 4 20 mA 20 0 mA 20 4 mA 0 10 V 2 10 V 10 0 V 10 2 V SCALING START 1 41 0 S cm 4 mA Can be set in the range being used depending on the...

Страница 39: ...VALUE 0 0 4 0 22 0 mA depending on the signal type 0 10 7 V MANUAL MODE OFF ON MAN VALUE 0 0 4 0 22 0 mA depending on the signal type 0 10 7 V 10 6 3 TEMPERATURE IN DIMENS UNIT C F MEAS MODE SENSOR M...

Страница 40: ...200 to 150 0 C 20 mA depending on the signal type DURING ALARM LOW 0 mA 0 V 3 4 mA 1 4 V HIGH 22 mA 10 7 V SAFE VALUE depending on the signal type DURING CALIBRATION MOVING FROZEN SAFE VALUE SAFE VALU...

Страница 41: ...DUCT MIN TEMP MAX TEMP LK1 TEMP LK2 TEMP CALIB TIMER ALARM MAX limit comparator MIN limit comparator LK1 alarm window LK2 alarm window Pulse contact Trigger condition longer than pulse duration Pulse...

Страница 42: ...ve SPACING 0 0 999 0 depending on the function see above MANUAL MODE OFF ON FOR HOLD INACTIVE ACTIVE FROZEN FOR ALARM CALIB INACTIVE ACTIVE FROZEN ON DELAY 00 00 00 01 00 00 H M S OFF DELAY 00 00 00 0...

Страница 43: ...8 00 00 H M S LOCK TIME 0 00 00 00 01 00 18 00 00 H M S Setting parameters Binary input 1 Binary input 2 Range temperature coefficient changeover Range 1 TC 1 open open Range 2 TC 2 closed open Range...

Страница 44: ...ne which parameters can be edited altered by a normal user and or which calibration actions are permitted Editable parameters can be edited in the operator level Non editable parameters are marked in...

Страница 45: ...e which parameters can be altered edited by the operator in the operator level The available options are READ ONLY and EDIT The structure Parameter level within the administrator level is identical to...

Страница 46: ...rating the relative cell constant Page 51 10 8 2 TEMPCO LINEAR linear temperature coefficient If this function has been enabled by the administrator then the operator can calibrate the device for liqu...

Страница 47: ...the dilution valve to be opened and a portion of the added biocide would be discharged into the waste water possibly contravening regulations To prevent this the conductivity in the cooling system is...

Страница 48: ...0 9 1 Stop dilution All the parameters are system dependent and must be adjusted to suit system requirements Press the key for at least 3 seconds Use the or key to select OPERATOR LEVEL use the key to...

Страница 49: ...Use the and or keys to set the dilution factor in the range from 1 10 50 below the limit value Confirm the selection with the key Use the or key to select DOSING TIME use the key to confirm the select...

Страница 50: ...and keys in the range from 0 00 00 00 01 00 18 00 00 H M S Confirm the setting with the key If there is an interruption in the supply voltage during dilution the function will be canceled The dilution...

Страница 51: ...applies to the split version The transmitter is in the measurement mode Immerse the conductivity sensor in a reference solution with a known conductivity Press the key for at least 3 seconds Use the...

Страница 52: ...value press the key The transmitter is in the calibration menu Press the key The transmitter is now in the measurement mode and shows the compensated conductivity of the reference solution 11 3 Calib...

Страница 53: ...the key for at least 3 seconds Use the and keys to select CALIBRATION LEVEL use the key to confirm the selection Use the and keys to select TEMPCO LINEAR use the key to confirm the selection Use the...

Страница 54: ...The transmitter is in the calibration menu Press the key The transmitter is now in the measurement mode and shows the compensated conductivity of the reference solution During calibration the rate of...

Страница 55: ...re range TStart to TEnd into 5 sections A different TC value can be used for compensation in each of these range sections This TC curve can be edited with the setup program and transmitted to the devi...

Страница 56: ...omatic calibration The TC curve is automatically recorded over a temperature range that has been defined by the user The temperature range between the start and end temperatures is subdivided into 5 s...

Страница 57: ...tivity sensor in a sample of the solution to be measured Press the key for at least 3 seconds Use the and keys to select CALIBRATION LEVEL use the key to confirm the selection Use the or key to select...

Страница 58: ...itself The LC display now shows at top 1 the next target temperature blinking below 2 the present sensor temperature steady Warm up the sample medium until is it above below the temperature that is b...

Страница 59: ...at least 3 seconds or to reject the values press the key The transmitter is in the calibration menu Press the key The transmitter is now in the measurement mode and shows the compensated conductivity...

Страница 60: ...sits on the sensor section can be removed with a soft brush e g a bottle brush Do not use solvents Hard to remove crusts and deposits can be softened and removed with dilute hydrochloric acid Observe...

Страница 61: ...able measurement display sensor not immersed deeply enough top up the reservoir inadequate mixing ensure good mixing for sensor all round free space of approx 5 mm ensure all round flow air bubbles ch...

Страница 62: ...alues above 50 mS the loop must have 3 turns Example 1 The measuring transmitter is to show 20 mS To achieve a display of 20 mS cm the resistance loop with 1 turn must have a resistance of 312 5 Ohm D...

Страница 63: ...e Chapter 10 6 1 CONDUCTIVITY IN Conductivity input Page 37 RANGE 1 4 Set TC to 0 C see Chapter 10 6 1 CONDUCTIVITY IN Conductivity input Page 37 TEMPCO Fit the resistance loop as shown in the diagram...

Страница 64: ...device see Chapter 7 Installation Page 25 Select the range appropriate to the conductivity test solution see Chapter 10 6 1 CONDUCTIVITY IN Conductivity input Page 37 RANGE 1 4 Set TC to 0 C see Chap...

Страница 65: ...N Conductivity input Page 37 TEMPCO Set the TC for the reference instrument to 0 C as well see operating instructions for the reference instrument If this is not possible then the sample liquid must b...

Страница 66: ...1000 S cm 0 2000 S cm 0 5000 S cm 0 10 mS cm 0 20 mS cm 0 50 mS cm 0 100 mS cm 0 200 mS cm 0 500 mS cm 0 1000 mS cm 0 2000 mS cm uncompensated 37 Temperature compensation linear non linear natural wa...

Страница 67: ...urement mode Sensor manual 39 Manual value 20 0 to 25 to 150 C 39 Offset 15 0 to 0 0 to 15 C 39 Filter time 00 00 00 00 00 01 00 00 25 H M S 39 Temperature output Signal type 0 20 mA 4 20 mA 20 0 mA 2...

Страница 68: ...s 0 0 0 5 999 0 42 Spacing 0 0 999 0 42 Manual mode off on 42 For hold inactive active frozen 42 For alarm calibration inactive active frozen 42 Switch on delay 00 00 00 01 00 00 H M S 42 Switch off d...

Страница 69: ...h Italian Dutch Polish Portuguese Russian Swedish Spanish 44 Contrast 0 6 11 44 Lighting off on during operation 44 LCD inverse off on 44 Parameter Selection value range Factory setting New adjustment...

Страница 70: ...0XWWHU 6FUHZ 6FKUDXEH d 7KLV WDEOH LV SUHSDUHG LQ DFFRUGDQFH ZLWK WKH SURYLVLRQV 6 7 7 QGLFDWH WKH KD DUGRXV VXEVWDQFHV LQ DOO KRPRJHQHRXV PDWHULDOV IRU WKH SDUW LV EHORZ WKH OLPLW RI WKH 7 7 QGLFDWH...

Страница 71: ...All manuals and user guides at all guides com a l l g u i d e s c o m...

Страница 72: ...many Phone 49 661 6003 0 Fax 49 661 6003 607 Email mail jumo net Internet www jumo net JUMO House Temple Bank Riverway Harlow Essex CM20 2DY UK Phone 44 1279 63 55 33 Fax 44 1279 62 50 29 Email sales...

Отзывы: