11 Calibration
58
✱
Use the
and
keys to enter the start temperature; confirm with the
key.
✱
Use the
and
keys to enter the end temperature; confirm with the
key.
The transmitter will define the fixed temperature points itself. The LC display
now shows
- at top (1): the next target temperature (blinking)
- below (2): the present sensor temperature (steady)
✱
Warm up the sample medium until is it above/below the temperature that is
blinking.
The start temperature must be lower than the reference
temperature (25.0°C).
The end temperature must be at least 20°C above the start
temperature.
PGM
PGM
(1)
(2)