18
6 Mounting
6.1
General
Mounting site
Make sure that the site is readily accessible, for calibration at a later time.
The fixing must be secure and free from vibration.
Avoid direct sunlight!
Take care that there is adequate flow through and around the sensor (1) !
If the device is to be mounted in a pipeline, there must be at least 20 mm
clearance between the sensor and the wall of the pipe.
If it is not possible to achieve this minimum clearance, then a limited
compensation can be made through the “Mounting factor” parameter.
For submerged operation in basins, a location must be chosen that is
representative of the typical conductivity or concentration.
Mounting
position
The measuring transmitter can be mounted in any position.
Screwing in and
unscrewing the
detached
sensor
The cable must not be twisted!
Avoid putting tension on the cable. In particular, avoid tugging it.
Sensor details
min.
20
mm
min.
20
mm
(1)