6. Maintenance & Check 60
6.6 Replacing Fuses
Exchange the fuse for the one of our specification. Exchange it after confirming the cause to which the fuse
is blown. Moreover, turn off the main switch of the power supply CQD-2083 when you exchange fuses
(Press
○
sign side).
No.
Model type
Rating
Remarks
F1
250V A TLC 5A
250V 5A
For power supply in this
equipment
F2
MF51NR 250V 0.5 or equivalent
250V 0.5A
For power supply alert
circuit in this equipment
F3
MF51NR 250V 2 or equivalent
250V 2A
For built-in printer power
supply
Fuse Positions
Inside of main unit
CQD-2083
(1) Replacing Main Power Supply Fuse F1
・
As for the cause, the defective cable connected with the power supply is considered.
Exchange fuses after defective checking the cable. Still, when the fuse is blown, the defect in the power
supply CBD-1811 is thought. Request the repair.
MAIN SW
○
side is OFF.
F1
F2
F3
Содержание JFE-680 -
Страница 2: ......
Страница 9: ...External View External View vii NO ALERT ...
Страница 18: ...1 5 System Configuration 1 Introduction 6 alert alert alert alert alert alert alert ...
Страница 46: ...4 Operation 34 2 HISTORY print mode 3 LOG print mode 24hours history 30minutes log 1hour time mark ...
Страница 54: ...5 1 Installing the Recorder Unit Flush Mount Equipment Figure 3 1 5 Installation 42 Unit mm ...
Страница 55: ...Wall Mount Equipment Figure 3 2 5 Installation 43 Unit mm ...
Страница 57: ...NKF 345 5 Installation 45 ...
Страница 58: ...NKF 392C 5 Installation 46 ...
Страница 59: ...NKF 393 5 Installation 47 ...
Страница 60: ...NKF 394 5 Installation 48 ...
Страница 61: ...NKF 396 5 Installation 49 ...
Страница 79: ...Appendix Noise Bubble Noise Bubble Interruption Interference Noise from other ship Plankton layer Appendix 67 ...
Страница 84: ......
Страница 85: ......