DINAMAP*
Compact
Vital Signs Monitor
42
Then carefully pull the top of the front panel away from the rear enclosure. The panel will tilt forwards
but is still retained by two plastic clips at its bottom edge. Unhook the front panel from these clips,
taking care not to damage them. This is particularly important when refitting the front panel later.
Before the panel can be separated from the case, it is necessary to disconnect the switch board cable,
temperature probe cable and SpO
2
sensor cable if fitted.
The Switch Board ribbon cable plugs into socket PL10 on the Main Board via a ferrite which clamps the
cable to the front of the chassis metalwork. Detach the ferrite from the chassis front plate by separating
the Dual-Lock
TM
fastener.
With the ferrite detached, the ribbon cable can be unplugged from socket PL10.