DINAMAP*
Compact
Vital Signs Monitor
11
1.7 Electrical, Mechanical & Environmental Specifications
NIBP
CUFF PRESSURE RANGE:
Adult / Paediatric: 0 mmHg to 290 mmHg
Neonate: 0 mmHg to 140 mmHg
DEFAULT TARGET :
Adult / Paediatric: 178 ± 15 mmHg
CUFF INFLATION
Neonate: 110 ± 15 mmHg
TARGET CUFF INFLATION :
Adult / Paediatric: 100 to 250 mmHg; 5 mmHg increments
ADJUSTMENT RANGE
Neonate: 100 to 140 mmHg; 5 mmHg increments
BLOOD PRESSURE
MEASUREMENT RANGE:
Systolic
mmHg
MAP mmHg
Diastolic
mmHg
Adult / Paediatric:
30 - 245
15 - 215
10 - 195
Neonate:
40 - 140
30 - 115
20 - 100
BLOOD PRESSURE ACCURACY:
Meets AAMI/ANSI SP-10 Electronic or Automated
Sphygmomanometer Standard for Accuracy. (AAMI/ANSI
standard: ± 5 mmHg mean error; 8 mmHg standard deviation;
intra-arterial method.)
BLOOD PRESSURE
20 seconds to 45 seconds typical
DETERMINATION TIME:
Adult: 120 seconds maximum,
Neonate: 85 seconds maximum.
PULSE RATE DETERMINATION:
Adult / Paediatric: 30 - 200 BPM
Neonate: 30 - 200 BPM
PULSE RATE ACCURACY:
± 3.5 percent
OVERPRESSURE CUT-OFF:
Adult / Paediatric 300 - 330 mmHg
Neonate 150 - 165 mmHg
Predictive Temperature Determination
RANGE:
Max. 42.2
°
Celsius 108.0° Fahrenheit
Min 31.1
°
Celsius 88.0° Fahrenheit
PROBE ACCURACY:
±
0.1
o
C ± 0.2
o
F
PREDICTIVE TEMPERATURE
±
0.6
o
C, 36.1
- 39.4
o
C ± 1
o
F, 97
- 103
o
F
ACCURACY :
UNSPECIFIED < 36.1
O
C, > 39.4
O
C
UNSPECIFIED < 97
O
F, > 103
O
F
DETERMINATION TIME:
30 seconds typical; 60 seconds maximum
SpO
2
(TS Model Only)
OXYGEN SATURATION
RANGE :
0 to 100%
SATURATION ACCURACY:
0% to 69% : unspecified
ADULT ACCURACY (70-100%)