DINAMAP*
Compact
Vital Signs Monitor
54
Select VSENSE from sub menu.
Select VAL2 from sub menu.
Increase applied pressure until overpressure failsafe occurs.
Confirm the following :-
Pressure at failsafe point is between 150mmHg to 165mmHg (200mBar to 220mBar)
The system pressure falls to less than 20mmHg (26.6mBar) within 4 seconds.
Switch off UUT.
If UUT fails re-calibrate unit as per section 3.3.4
4.2.5.3.2 Adult Overpressure Limit Check
Switch on UUT.
Enter Service mode as detailed in section 4.1.3.1
Select NIBP and then CHECK.
Increase applied pressure until overpressure failsafe occurs.
Confirm the following :-
Pressure at failsafe point is between 300mmHg to 330mmHg (400mBar to 440mBar)
The system pressure falls to less than 20mmHg (26.6mBar) within 4 seconds.
Switch off UUT and disconnect calibration kit.
If UUT fails re-calibrate unit as per section 3.3.4
4.2.6 Printer Operation
Switch on UUT.
The operation of the internal printer shall be tested if fitted, otherwise ignore this section.
If printer is fitted and PRINT option is unavailable on main operational menu of the UUT LCD.
Enter Service Mode as detailed in section 4.1.3.1
Select SYSTEM from sub menu.
Select SYSTEM from sub menu again.
Select PRINTER option box on UUT LCD using rotor switch, press rotor push button and confirm it has
a tick beside it.
Switch off UUT and ensure a paper roll is fitted in the printer and the printer door is shut.
4.2.6.1 Print Quality
Confirm printer status on LCD indicates ‘Man’.
Select PRINT option from main menu followed by NOW. Check printer output and confirm that all
characters are readable and have an even contrast.
4.2.6.2 Printer Error
Open printer cover .
From the main menu select PRINT and NOW.
Confirm that the printer does not start and LCD printer status display indicates ‘No Paper’..
Remove paper and close printer cover.
Confirm that the printer does not start.
Replace paper and close cover.
Select PRINT and NOW and confirm printer starts.
4.2.7 Calibration Time Stamp
If all temperature readings correct switch off UUT and switch on again.
Enter Service Mode as detailed in section 4.1.3.1