Joerns® Bed Frames

WeCare™ Bed

© 2022 Joerns Healthcare • 6110435 RevC • 22-2083

Warning: Possible Injury Or Death

. Do not use an assist 

device if any openings within assist device body allow 

resident/patient to get his/her head or neck lodged within 

these openings. Use of assist devices with large openings, 

may result in injury or death.

Warning: Possible Injury Or Death.

 Do not use any 

Joerns

®

 assist device until you verify it is designed 

specifically for the WeCare bed and it is locked in place. 

Failure to lock assist devices may result in injury or death.

Warning: Possible Injury or Death. 

NEVER permit more 

than (1) person in /on the bed at any time. NEVER exceed 

the weight capacity of this bed. The weight capacity of this 

bed is 450 lbs (204 kg) including accessories and options 

(not all accessories and options may be used at the same 

time). Subtract residents/patients weight from beds capacity 

450 lbs (204 kg) to determine allowable weight for mattress, 

bed panels, accessories, and optional equipment. See 

Accessories Section for accessory weights.

Warning: Possible Injury Or Death

. Do not use any 

replacement parts not manufactured, marketed, or provided 

by Joerns Healthcare on any Joerns bed. Use of unapproved 

replacement parts may result in injury or death.

Warning: Possible Injury Or Death.

 If using accessories 

not manufactured, marketed, or provided by Joerns 

Healthcare for Joerns beds, consult with the manufacturer 

for compatibility and limitations prior to use. Failure to do so 

may result in injury or death.

Caution:

 Radio transmitting equipment, cellular phones, 

or similar electronic devices, used in proximity of bed, may 

affect bed performance.

Caution:

 Particular precaution must be considered 

during exposure to strong emission sources such as High 

Frequency surgical equipment and  similar devices. Do not 

route High Frequency cables on or near device. If in doubt, 

contact a qualified technician.

Note:

 Electromagnetic disturbances may contribute to the 

improper operation of medical devices.  In the case of a bed, 

it may affect the performance of the bed movement when a 

function is selected.  Do not place the bed in a location that 

is known to have high electromagnetic fields.

Warning:

 Use of this equipment adjacent to or stacked with 

other equipment should be avoided because it could result in 

improper operation. If such use is necessary, this equipment 

and the other equipment should be observed to verify that 

they are operating normally.

Warning:

 Use of accessories, transducers, and cables 

other than those specified or provided by the manufacturer 

of this equipment could result in increased electromagnetic 

emissions or decreased electromagnetic immunity of this 

equipment and result in improper operation.

Warning: 

Portable RF communications equipment (including 

peripherals such as antenna cables and external antennas) 

should be used no closer than 30 cm (12 inches) to any 

part of the WeCare bed, including cables specified by the 

manufacturer. Otherwise, degradation of the performance of 

this equipment could result.

Note:

 The EMISSIONS characteristics of this equipment 

make it suitable for use in industrial areas and hospitals 

(CISPR 11 class A). If it is used in a residential environment 

(for which CISPR 11 class B is normally required) this 

equipment might not offer adequate protection to radio-

frequency communication services. The user might need to 

take mitigation measures, such as relocating or re-orienting 

the equipment.

Caution:

 This device complies with EMC requirements 

of IEC 60601-1-2. Radio transmitting equipment, cellular 

phones or similar electronic devices, used in proximity of 

bed, may affect bed performance.

Caution:

 In the absence of line voltage, all bed adjustments 

can be achieved via an externally available backup battery 

(Joerns P/N N211).

Warning: Possible Shock Hazard.

 If the bed power cord has 

a hospital-grade, 3-prong grounding plug, grounding reliability 

can only be achieved when equipment is connected to an 

equivalent receptacle marked hospital-grade.

Warning: Possible Injury.

 Sitting on or sliding over the 

Mattress Side Stop can cause skin injury if the bed mattress 

compresses to less than 1.5 inches under user’s weight.

Warning: Possible Injury. 

Risk of entanglement or injury 

may occur if mattress used with Mattress Side Stop does not 

fill the entire width between stops or which compresses to 

less than 1.5 inches under user’s weight.

Caution: 

The Joerns’ WeCare bed is intended for use within 

institutional healthcare environments 2, 3, and 5 (ie: skilled 

nursing, transitional care, rehabilitational care, assisted living). 

Joerns Healthcare recommends compliance to the regulations 

and guidelines specific to your locality.

Joerns assist devices are designed to be used with specific 

Joerns bed frames. Once mounted on the appropriate bed 

frame, assist devices become an integral part of the Joerns 

bed system. As such, these mounted assist devices, meet the 

IEC 60601-2-52 Bed Standard, within normal use conditions, 

of the appropriate bed to which they are mounted. 

Assemble, install, utilize, and maintain Joerns Assist Devices 

as described in their accompanying Joerns User/ Service 

Manual, which also lists the bed type or model with which 

these devices may be used. 

This product is only one part of your healthcare bed 

system. Proper combinations of mattress, head/foot panels 

and assist devices are needed to minimize the risk of 

entrapment. For more information, call Customer Care, 800-

826-0270.

There are no known contraindications for this device.

Содержание WeCare Bed

Страница 1: ...User Service Manual Joerns Bed Frames WeCare Bed To avoid injury read user s manual before using...

Страница 2: ...actory installed routing configuration Failure to do so could result in injury or death Warning An optimal bed system assessment should be conducted on each resident patient by a qualified clinician o...

Страница 3: ...lectromagnetic immunity of this equipment and result in improper operation Warning Portable RF communications equipment including peripherals such as antenna cables and external antennas should be use...

Страница 4: ...r of long term care beds and a frontrunner in addressing this critical issue Joerns Healthcare can offer you the expertise assistance and products to bring your facility into compliance Joerns Complia...

Страница 5: ...ondensing Atmospheric Pressure 700 hPa to 1060 hPa Protected Against Splashing Water IPX4 Storage and Shipping Conditions Ambient Temperature 40 C to 70 C Relative Humidity 10 to 100 Atmospheric Press...

Страница 6: ...tting orange zip ties 1 Identify head frame and foot frame by compatible ID labels on left side Figure 1 2 Lay frame halves on soft surface feet side down Place packing material under head frame to ma...

Страница 7: ...and head end hi lo cables and route through safety pin bail Figure 6 then route behind cable bundle Figure 10 and 11 8 Insert back actuator cable in port 1 and insert hi lo actuator head end cable in...

Страница 8: ...ols Required 2 Phillips screwdriver 1 Locate the screws at each corner of the inside tube at the foot end of the screws mattress support platform one per side Figure 16 2 Remove the screws using the 2...

Страница 9: ...will have staff control electronics if staff control was ordered 2 Lay panel on soft surface front face down 3 Foot panel brackets are shorter of two sizes Figure 19 4 Install the foot panel brackets...

Страница 10: ...end 3 Unplug the pendant Route one cable under the control box and plug into pendant port Route the second cable under the control box and connect to the hand pendant Figure 22 4 Route the two cables...

Страница 11: ...art with the bed in the highest position Press Trendelenburg Reverse Trendelenburg button until desired angle is obtained Caution Ensure the bed remains stable and does not roll sideways while in Tren...

Страница 12: ...Possible high temperature Prior to servicing any electrical component ensure the component has cooled to room temperature Power Cord Conduct a yearly inspection of power cord to ensure efficient bed o...

Страница 13: ...ng Always follow proper factory installed routing configuration Be sure to check condition of power cord on a regular basis Do not use damaged or frayed power cords 4 Disconnect the ground wire if pre...

Страница 14: ...of cable Avoid dropping or stretching pendant during use to maximize service life Cleaning Caution Possible damage to electrical parts and or shock hazard Warning Possible Injury or Burns Do not use f...

Страница 15: ...crossing over thresholds may cause it to disconnect from bed resulting in loss of control of bed 4 Take extra care when moving bed remember do not move bed while occupied by resident patient on any sl...

Страница 16: ...loor where they might be stepped on and adversely affected or damaged Exercise particular caution when the bed is moved to its lowest position as items attached to the holder at that height are especi...

Страница 17: ...cations equipment and the Joerns WECC Bed The Joerns WECC bed is intended for use in an electromagnetic environment for professional healthcare in which radiated RF disturbances are controlled The cus...

Страница 18: ...al healthcare environment Surge IEC 61000 4 5 0 5 kV 1 kV line s to line s 0 5 kV 1 kV 2 kV line s to earth 0 5 kV 1 kV line s to line s 0 5 kV 1 kV 2 kV line s to earth Mains power quality should be...

Страница 19: ...MHz 80 MHz 80 AM at 1 kHz 3 V m 80 MHz 2 7 GHz 80 AM at 1 kHz Portable and mobile RF communications equipment should be used no closer to any part of the Joerns WECC bed including cables than the reco...

Страница 20: ...re will void this warranty Parts Joerns beds contain various parts that wear from normal use These parts such as DC batteries are not covered under the one year warranty but are covered for 90 days af...