Actuator and Control Box Troubleshooting

Any loss of electrical function should be investigated 

with the following steps:

1. 

Make sure that all cables are plugged in, including 

pendant cable and AC power cord.

2.  Switch pendant with a new one or one from a bed 

that is functioning normally.

3. 

If these steps fail to resolve the problem, call 

Technical Support at 800-826-0270.

To Remove Actuator

Recommended Tools: Screwdriver, Hammer, Punch

1. 

Identify actuator to be replaced.

2.  Tip bed on its side or upside down to replace  

hi-lo actuators.

3.  Mark cable ties locations with tape for later 

reference. Cut cable ties that hold the motor 

cables to frame.

4. 

Actuator is held in place by two pins. Use a 

screwdriver to remove clip on each pin.

5. 

Slide pins out of holes. If necessary, use hammer 

and punch to tap out the pins.

6. 

You may now replace the actuator(s).

7.  To reassemble bed, reverse the previous steps, 

making note of the following two important items:

• 

Assemble all hardware as originally installed.

8. 

Install cable ties per designated locations marked 

in step 3 by tape.

To Remove AC Power Cord

Recommended Tools: Small Screwdriver, Masking 

Tape, Cable Ties

1.  Raise bed to maximum height if possible.

2.  Unplug power cord from wall.

3. 

Cut cable ties that hold AC power cord to frame. 

Tape remaining cables in their original position. It 

is very important to keep actuator cables routed 

along frame in original factory-installed positions. 

If this step is not followed, damage to cables or 

electric shock may result.

Warning: Possible Shock Hazard.

 Injury may result 

from improper power cord routing. Always follow 

proper factory-installed routing configuration. Be sure 

to check condition of power cord on a regular basis. 

Do not use damaged or frayed power cords.

 

4. 

Disconnect the ground wire, if present.

5. 

Using a small screwdriver, alternately depress 

small latch tabs above and below AC inlet 

receptacle on the control box. Simultaneously, 

pull gently outward on AC power cord where it 

connects to control box until plug comes out.

6.  Plug replacement cord into control box until  

it latches. 

7.  Re-route AC power cord along frame in same 

position as original routing.

8. 

Replace cable ties along with cables in their 

original position and routing to the control box. 

If this step is not followed, damage to cables or 

electric shock may result.

9.  Ensure ground wire is re-attached, if applicable.

Troubleshooting

Any loss of electrical function should be investigated 

with the following steps:

Event

Possible Remedies

Bed not 

responding to 

pendant control

1. 

Check all connections are fully 

inserted with cable secured. Check 

cables for damage or fraying.

2. 

Check actuator pins are connected 

and secure.

3. 

Check the bed is plugged into 

power source and control box green 

light is illuminated

4.  Thermal overload condition. This 

should self-correct in 2 minutes 

or less. Do not surpass 10% duty 

cycle (2 min on/ 18 min off).

5. 

Check if bed is not overloaded and 

clear of obstructions

6.  If these steps fail to resolve the 

problem, call Technical Support at 

800-826-0270.

Back sleep 

surface or Knee 

surface does not 

raise or lower

1.  Ensure the actuator pins and 

clips were installed properly 

during assembly

2. 

Check the actuator cables are 

fully inserted into control box in 

the correct port location

Bed moves 

across the floor 

at low height

1. 

Check caster locks are engaged

Note:

 Electrical damage can occur if a non-approved 

pendant is connected to the electrical system.

Содержание WeCare Bed

Страница 1: ...User Service Manual Joerns Bed Frames WeCare Bed To avoid injury read user s manual before using...

Страница 2: ...actory installed routing configuration Failure to do so could result in injury or death Warning An optimal bed system assessment should be conducted on each resident patient by a qualified clinician o...

Страница 3: ...lectromagnetic immunity of this equipment and result in improper operation Warning Portable RF communications equipment including peripherals such as antenna cables and external antennas should be use...

Страница 4: ...r of long term care beds and a frontrunner in addressing this critical issue Joerns Healthcare can offer you the expertise assistance and products to bring your facility into compliance Joerns Complia...

Страница 5: ...ondensing Atmospheric Pressure 700 hPa to 1060 hPa Protected Against Splashing Water IPX4 Storage and Shipping Conditions Ambient Temperature 40 C to 70 C Relative Humidity 10 to 100 Atmospheric Press...

Страница 6: ...tting orange zip ties 1 Identify head frame and foot frame by compatible ID labels on left side Figure 1 2 Lay frame halves on soft surface feet side down Place packing material under head frame to ma...

Страница 7: ...and head end hi lo cables and route through safety pin bail Figure 6 then route behind cable bundle Figure 10 and 11 8 Insert back actuator cable in port 1 and insert hi lo actuator head end cable in...

Страница 8: ...ols Required 2 Phillips screwdriver 1 Locate the screws at each corner of the inside tube at the foot end of the screws mattress support platform one per side Figure 16 2 Remove the screws using the 2...

Страница 9: ...will have staff control electronics if staff control was ordered 2 Lay panel on soft surface front face down 3 Foot panel brackets are shorter of two sizes Figure 19 4 Install the foot panel brackets...

Страница 10: ...end 3 Unplug the pendant Route one cable under the control box and plug into pendant port Route the second cable under the control box and connect to the hand pendant Figure 22 4 Route the two cables...

Страница 11: ...art with the bed in the highest position Press Trendelenburg Reverse Trendelenburg button until desired angle is obtained Caution Ensure the bed remains stable and does not roll sideways while in Tren...

Страница 12: ...Possible high temperature Prior to servicing any electrical component ensure the component has cooled to room temperature Power Cord Conduct a yearly inspection of power cord to ensure efficient bed o...

Страница 13: ...ng Always follow proper factory installed routing configuration Be sure to check condition of power cord on a regular basis Do not use damaged or frayed power cords 4 Disconnect the ground wire if pre...

Страница 14: ...of cable Avoid dropping or stretching pendant during use to maximize service life Cleaning Caution Possible damage to electrical parts and or shock hazard Warning Possible Injury or Burns Do not use f...

Страница 15: ...crossing over thresholds may cause it to disconnect from bed resulting in loss of control of bed 4 Take extra care when moving bed remember do not move bed while occupied by resident patient on any sl...

Страница 16: ...loor where they might be stepped on and adversely affected or damaged Exercise particular caution when the bed is moved to its lowest position as items attached to the holder at that height are especi...

Страница 17: ...cations equipment and the Joerns WECC Bed The Joerns WECC bed is intended for use in an electromagnetic environment for professional healthcare in which radiated RF disturbances are controlled The cus...

Страница 18: ...al healthcare environment Surge IEC 61000 4 5 0 5 kV 1 kV line s to line s 0 5 kV 1 kV 2 kV line s to earth 0 5 kV 1 kV line s to line s 0 5 kV 1 kV 2 kV line s to earth Mains power quality should be...

Страница 19: ...MHz 80 MHz 80 AM at 1 kHz 3 V m 80 MHz 2 7 GHz 80 AM at 1 kHz Portable and mobile RF communications equipment should be used no closer to any part of the Joerns WECC bed including cables than the reco...

Страница 20: ...re will void this warranty Parts Joerns beds contain various parts that wear from normal use These parts such as DC batteries are not covered under the one year warranty but are covered for 90 days af...

Отзывы: