Appendix A: Electromagnetic Compatibility (EMC) Information

Manufacturer’s Declaration –Electromagnetic Emissions

The Joerns WECC bed is intended for use in the electromagnetic environment (for professional healthcare) 

specified below. The customer or the user of the Joerns WECC bed should assure that it is used in such an 

environment.

Emissions test

Compliance

Electromagnetic environment-guidance (for professional 

healthcare environment)

RF emissions 

CISPR 11

Group 1

The Joerns WECC bed uses RF energy only for its internal function. 

Therefore, its RF emissions are very low and are not likely to cause 

any interference in nearby electronic equipment.

RF emissions 

CISPR 11

Class A

The Joerns WECC bed is suitable for use in all establishments other 

than domestic and those directly connected to the public low-voltage 

power supply network that supplies buildings used for domestic 

purposes.

Harmonic emissions

IEC 61000-3-2

Class A

Voltage fluctuations / 

 

flicker emissions 

IEC 61000-3-3

Compliance

Recommended separation distance between portable and  

mobile RF communications equipment and the Joerns WECC Bed

The Joerns WECC bed is intended for use in an electromagnetic environment (for professional healthcare) in which 

radiated RF disturbances are controlled. The customer or the user of the Joerns WECC bed can help prevent 

electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications 

equipment (transmitters) and the Joerns WECC bed as recommended below, according to the maximum output 

power of the communications equipment.

Rated maximum output 

power of transmitter

W

Separation distance according to frequency of transmitter

 m

150 kHz to 80 MHz

d=1,2√P

80 MHz to 800 MHz

d=1,2√P

800 MHz to 2,7 GHz

d=2,3√P

0,01

0,12

0,12

0,23

0,1

0,38

0,38

0,73

1

1,2

1,2

2,3

10

3,8

3,8

7,3

100

12

12

23

For transmitters rated at a maximum output power not listed above, the recommended separation distance 

d

 in 

meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where 

P

 is the 

maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer. 

Note 1: 

At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.

Note 2: 

These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and 

reflection from structures, objects, and people.

Содержание WeCare Bed

Страница 1: ...User Service Manual Joerns Bed Frames WeCare Bed To avoid injury read user s manual before using...

Страница 2: ...actory installed routing configuration Failure to do so could result in injury or death Warning An optimal bed system assessment should be conducted on each resident patient by a qualified clinician o...

Страница 3: ...lectromagnetic immunity of this equipment and result in improper operation Warning Portable RF communications equipment including peripherals such as antenna cables and external antennas should be use...

Страница 4: ...r of long term care beds and a frontrunner in addressing this critical issue Joerns Healthcare can offer you the expertise assistance and products to bring your facility into compliance Joerns Complia...

Страница 5: ...ondensing Atmospheric Pressure 700 hPa to 1060 hPa Protected Against Splashing Water IPX4 Storage and Shipping Conditions Ambient Temperature 40 C to 70 C Relative Humidity 10 to 100 Atmospheric Press...

Страница 6: ...tting orange zip ties 1 Identify head frame and foot frame by compatible ID labels on left side Figure 1 2 Lay frame halves on soft surface feet side down Place packing material under head frame to ma...

Страница 7: ...and head end hi lo cables and route through safety pin bail Figure 6 then route behind cable bundle Figure 10 and 11 8 Insert back actuator cable in port 1 and insert hi lo actuator head end cable in...

Страница 8: ...ols Required 2 Phillips screwdriver 1 Locate the screws at each corner of the inside tube at the foot end of the screws mattress support platform one per side Figure 16 2 Remove the screws using the 2...

Страница 9: ...will have staff control electronics if staff control was ordered 2 Lay panel on soft surface front face down 3 Foot panel brackets are shorter of two sizes Figure 19 4 Install the foot panel brackets...

Страница 10: ...end 3 Unplug the pendant Route one cable under the control box and plug into pendant port Route the second cable under the control box and connect to the hand pendant Figure 22 4 Route the two cables...

Страница 11: ...art with the bed in the highest position Press Trendelenburg Reverse Trendelenburg button until desired angle is obtained Caution Ensure the bed remains stable and does not roll sideways while in Tren...

Страница 12: ...Possible high temperature Prior to servicing any electrical component ensure the component has cooled to room temperature Power Cord Conduct a yearly inspection of power cord to ensure efficient bed o...

Страница 13: ...ng Always follow proper factory installed routing configuration Be sure to check condition of power cord on a regular basis Do not use damaged or frayed power cords 4 Disconnect the ground wire if pre...

Страница 14: ...of cable Avoid dropping or stretching pendant during use to maximize service life Cleaning Caution Possible damage to electrical parts and or shock hazard Warning Possible Injury or Burns Do not use f...

Страница 15: ...crossing over thresholds may cause it to disconnect from bed resulting in loss of control of bed 4 Take extra care when moving bed remember do not move bed while occupied by resident patient on any sl...

Страница 16: ...loor where they might be stepped on and adversely affected or damaged Exercise particular caution when the bed is moved to its lowest position as items attached to the holder at that height are especi...

Страница 17: ...cations equipment and the Joerns WECC Bed The Joerns WECC bed is intended for use in an electromagnetic environment for professional healthcare in which radiated RF disturbances are controlled The cus...

Страница 18: ...al healthcare environment Surge IEC 61000 4 5 0 5 kV 1 kV line s to line s 0 5 kV 1 kV 2 kV line s to earth 0 5 kV 1 kV line s to line s 0 5 kV 1 kV 2 kV line s to earth Mains power quality should be...

Страница 19: ...MHz 80 MHz 80 AM at 1 kHz 3 V m 80 MHz 2 7 GHz 80 AM at 1 kHz Portable and mobile RF communications equipment should be used no closer to any part of the Joerns WECC bed including cables than the reco...

Страница 20: ...re will void this warranty Parts Joerns beds contain various parts that wear from normal use These parts such as DC batteries are not covered under the one year warranty but are covered for 90 days af...

Отзывы: