Important Precautions

Warning: Possible Injury Or Death.

 This product is 

intended for use as an adjustable mattress platform 

that contours for resident/patient comfort with 

adjustable height for resident/patient and caregiver 

convenience. Use of this product in a manner for 

which it was not designed could result in unproven  

or unsafe configuration, potentially resulting in injury 

 

or death.

Warning: Possible Injury. 

Before adjusting bed, 

ensure that area under and near bed perimeter is free 

of people and obstructions. Failure to do so could 

result in injury or damage to the bed.

Warning: Possible Injury. 

Keep bed in lowest 

position except when providing care (bathing, clothing 

changes, etc.). Bed should be at lowest convenient 

height for entry or exit. Failure to do so could result in 

injury.

Warning: Possible Injury Or Death.

 Resident/

patients may become entangled in pendant cord. 

Resident/patients with reduced mental acuity should 

not be allowed access to pendant. Unsupervised use 

of pendant could result in injury or death.

Warning: Possible Injury Or Death.

 Bed safe 

working load is 500 pounds. This is total weight 

counting resident/patient, mattress, bedding 

accessories and any other equipment or persons likely 

to be on bed. Do not exceed 500-pound safe working 

load. Exceeding the safe working load could result in 

property damage, injury or death.

Warning: Possible Injury Or Death. 

Do not roll bed 

over abrupt thresholds with resident/patient in bed. 

Bed is not designed for transporting resident/patient. 

Transporting resident/patient in the bed could result in 

injury or death.

Warning: 

Possible fire hazard when used with oxygen 

administrating equipment other than nasal, mask or ½ 

bed length tent type. Not for use with full bed length 

tent or hyperbaric oxygen chamber. Oxygen tent 

should not extend below mattress support platform. 

Lock hand control at foot of bed when using oxygen 

administering equipment.

Warning: Possible Shock Hazard.

 Unplug 

power cord from wall outlet before performing any 

maintenance, cleaning or service to the bed. Failure to 

do so could result in injury or death. 

Note:

 In case of 

unexpected movement, unplug bed from any outlet.

Warning: Possible Shock Hazard.

 Injury may result 

from improper routing of the power cord. Always follow 

the proper factory-installed routing configuration. 

Failure to do so could result in injury or death.

Warning: Possible Injury Or Death. 

Use a mattress 

that is properly sized to fit mattress deck, that will 

remain centered on mattress deck relative to State and 

Federal guidelines. Joerns Healthcare recommends 

the use of a mattress with minimum dimensions of 35 

inches wide and 6 inches deep. Length should match 

mattress support platform. Use of an improperly fitted 

mattress could result in injury or death. If using the 

CareWide accessory or the bed is equipped with Care-

Fit integrated width expansion, Joerns Healthcare 

recommends the use of a mattress with a minimum 

width of 39 inches for a 39 inch setting and 42 inches 

for 42 inch setting.

Warning: Possible Injury Or Death.

 If a resident/

patient’s mental or physical condition could lead to 

resident/patient entrapment, the mattress support 

platform should be left in the flat position when 

unattended. Failure to do so could result in injury or 

death.

Warning:

 An optimal bed system assessment should 

be conducted on each resident by a qualified clinician 

or medical provider to ensure maximum safety of 

the resident. The assessment should be conducted 

within the context of, and in compliance with, the 

state and federal guidelines related to the use of 

restraints and bed system entrapment guidance, 

including the 

Clinical Guidance for the Assessment 

and Implementation of Side Rails 

published by the 

Hospital Bed Safety Workgroup of the U.S. Food 

and Drug Administration.

Warning: Possible Injury Or Death.

 Use a properly 

sized mattress in order to minimize the gap between 

the side of mattress and assist device. This gap must 

be small enough to prevent resident/patient from 

getting his/her head or neck caught in this location. 

Make sure that raising or lowering bed, or contouring 

the sleep surface, does not create any hazardous 

gaps. Excessive gaps may result in injury or death. 

Warning: Possible Injury Or Death. 

Locate mounting 

brackets for assist devices strictly according to 

instructions. Gap between head/foot panel and assist 

device must be small enough or large enough to 

prevent resident/patient from getting his/her head or 

neck caught in this location (see specific Assist Device 

Installation Instructions for more information). 

 

Содержание EasyCare

Страница 1: ...User Service Manual Joerns Bed Frames EasyCare Bed To avoid injury read user s manual before using...

Страница 2: ...rom any outlet Warning Possible Shock Hazard Injury may result from improper routing of the power cord Always follow the proper factory installed routing configuration Failure to do so could result in...

Страница 3: ...if mattress used with Mattress Side Stop does not fill the entire width between stops or which compresses to less than 1 5 inches under user s weight Warning Possible Injury or Death Floor locks incr...

Страница 4: ...Therefore compliance is the responsibility of the facility As the leading manufacturer of long term care beds and a frontrunner in addressing this critical issue Joerns Healthcare can offer you the ex...

Страница 5: ...tection LA27 Actuators Auto Reset Current Sensor Environmental Conditions Operating Conditions Ambient Temperature 10 C to 40 C Relative Humidity 30 to 75 Non Condensing Atmospheric Pressure 700 hPa t...

Страница 6: ...ain support of the bed panel Insert the safety pins on each side and latch them Figure 3 6 Press the mounting bracket against the bed frame and tighten the four panel mounting screws right and left si...

Страница 7: ...h 12 printed on it It is longer than the standard bumper 1 Remove the shorter bumper by tapping the tangs on either side of the leg with a rubber mallet 2 Place the 12 baseboard bumper 3 Tap into plac...

Страница 8: ...3 Ensure that the cable plug is fully inserted into the cable jack 4 To secure tighten the union nut on both cables Figure 14 Operation Pendant Operation Your EasyCare bed comes standard with one of t...

Страница 9: ...tion button Standard Staff Control Figure 21 2 1 4 6 5 3 7 Advanced Staff Control Figure 22 8 2 12 7 1 4 6 5 3 9 10 11 Head Operation 1 Push to raise Back 2 Push to lower Back Knee Operation 3 Push to...

Страница 10: ...Install correct baseboard bumper Bed Transport Roll in Low Lock System 1 To keep the bed from rolling when in the lowest position press down on the locking lever to lock casters Figure 23 2 The stand...

Страница 11: ...re 28 and 29 5 Release the tether to engage the detent Figure 27 6 Repeat the above steps on the remaining MSP extensions Warning Possible Injury Or Death Use a mattress that is properly sized to fit...

Страница 12: ...in intact and are securely fastened to their intended connection point Actuator and Control Box Information Warning Possible Shock Hazard Unplug the power cord from the wall outlet before performing a...

Страница 13: ...iver Masking Tape Zip Ties 5 16 Nut Driver 1 Raise the bed so that there is about 27 inches of clearance between the floor and the mattress support platform of bed 2 Unplug the power cable from the ou...

Страница 14: ...k of use check all threaded fasteners for looseness and make sure all pins are in their normal location and fastened securely Check monthly for loose bolts nuts pins and other retaining hardware Tight...

Страница 15: ...the floor and or cause the bed to become unstable resulting in injury or death Drainage Bag Holder Warning Possible BioHazard Prior to use consult the Drain Bag manufacturer s installation and use in...

Страница 16: ...igure 36 1 The Drainage Bag Holder is designed as an attachment point for items such as urine collection bags Make sure items attached to holder are safely secured Prior to use please read important w...

Страница 17: ...be fully charged Note We recommend that users charge up the Battery immediately upon receipt 4 The Battery has a useful life of about 3 5 years depending upon usage It is recommended to discharge the...

Страница 18: ...are covered for 90 days after date of delivery Joerns Healthcare s obligation under this warranty is limited to supplying replacement parts servicing or replacing at its option any product which is fo...