Setting 12 Degree Trendelenburg Mode 

(Advanced Staff Control)

The bed will ship set to 4° Gravity Assist mode. This 

mode limits the head down angle to 4°. The software 

can be changed to 12° Trendelenburg mode to enable 

Trendelenburg functionality.

1.  Install Advanced Staff Control Panel described.

2. 

Simultaneously, press and hold Trendelenburg 

(Button 10, Figure 22 on page 10) and reverse 

Trendelenburg (Button 11, Figure 22, page 10) 

buttons for 5 seconds on Advanced Staff Control 

Panel.

3.  Trendelenburg/Reverse Trendelenburg lock icon 

will begin flashing indicating that the system is in 

programming mode.

4. 

Press bed up (Figure 22, page 10) button to select 

full 12° Trendelenburg mode.

5. 

Bed Hi-Low lock out icon will flash twice confirming 

software is set to 12° Trendelenburg mode.  

Note: 

Be sure correct baseboard bumper is 

installed. Failure to install the correct baseboard 

bumper can result in damage to the bed or walls. 

The 12° bumper is identified by a silver label with 

“12°” printed on it.

6. 

To return bed to 4° Gravity Assist, repeat steps 2 

and 3 then press bed down (Figure 22, page 10). 

Bed Hi-Low lock icon will blink once confirming 

software is now in 4° Gravity Assist mode. Install 

correct baseboard bumper. 

Bed Transport

Roll in Low Lock System

1.  To keep the bed from rolling when in the lowest 

position, press down on the locking lever to lock 

casters (Figure 23).

2.  The standard caster set comes with two locking 

casters placed at opposing corners of bed.

Mobile
Position

Locked  
Position

Swivel 
Casters

Roll in 

Low Lock 

System

Low only Any Height 

(casters must 

be locked in 

Low)

Head & 

Foot (in Low 

position only)

Figure 23

Care-Lock Feature (Optional)

The bed features a rolling base that allows it to be  

rolled or locked at any bed height.  

To stop motion of the bed, enable the Care-Lock 

feature:

1. 

Engage the steering locks, located on the dual 

 

wheel casters at the head of the bed (Figure 24). 

2.  Engage the Care-Lock at the foot end of the bed 

by pressing down on the red or right side pedal 

(Figure 25).

To resume bed movement, disable the Care-Lock 

feature:

1. 

Disengage by pressing down on the green or left 

pedal (Figure 26). The steering locks may remain 

in place for desired maneuverability.

Warning: Possible Injury

. Before leaving the bed 

unattended, check that the Care-Lock feature is in 

the locked position. Never leave an unlocked bed 

unattended. 

Warning: Possible Injury.

 Check that the Care-Lock 

feature is in the locked position before allowing the 

resident to enter or exit the bed. 

For best practices, the Care-Lock feature should 

 

be locked at all times, except when the bed is 

being moved. 

Figure 24

Figure 25

Figure 26

Содержание EasyCare

Страница 1: ...User Service Manual Joerns Bed Frames EasyCare Bed To avoid injury read user s manual before using...

Страница 2: ...rom any outlet Warning Possible Shock Hazard Injury may result from improper routing of the power cord Always follow the proper factory installed routing configuration Failure to do so could result in...

Страница 3: ...if mattress used with Mattress Side Stop does not fill the entire width between stops or which compresses to less than 1 5 inches under user s weight Warning Possible Injury or Death Floor locks incr...

Страница 4: ...Therefore compliance is the responsibility of the facility As the leading manufacturer of long term care beds and a frontrunner in addressing this critical issue Joerns Healthcare can offer you the ex...

Страница 5: ...tection LA27 Actuators Auto Reset Current Sensor Environmental Conditions Operating Conditions Ambient Temperature 10 C to 40 C Relative Humidity 30 to 75 Non Condensing Atmospheric Pressure 700 hPa t...

Страница 6: ...ain support of the bed panel Insert the safety pins on each side and latch them Figure 3 6 Press the mounting bracket against the bed frame and tighten the four panel mounting screws right and left si...

Страница 7: ...h 12 printed on it It is longer than the standard bumper 1 Remove the shorter bumper by tapping the tangs on either side of the leg with a rubber mallet 2 Place the 12 baseboard bumper 3 Tap into plac...

Страница 8: ...3 Ensure that the cable plug is fully inserted into the cable jack 4 To secure tighten the union nut on both cables Figure 14 Operation Pendant Operation Your EasyCare bed comes standard with one of t...

Страница 9: ...tion button Standard Staff Control Figure 21 2 1 4 6 5 3 7 Advanced Staff Control Figure 22 8 2 12 7 1 4 6 5 3 9 10 11 Head Operation 1 Push to raise Back 2 Push to lower Back Knee Operation 3 Push to...

Страница 10: ...Install correct baseboard bumper Bed Transport Roll in Low Lock System 1 To keep the bed from rolling when in the lowest position press down on the locking lever to lock casters Figure 23 2 The stand...

Страница 11: ...re 28 and 29 5 Release the tether to engage the detent Figure 27 6 Repeat the above steps on the remaining MSP extensions Warning Possible Injury Or Death Use a mattress that is properly sized to fit...

Страница 12: ...in intact and are securely fastened to their intended connection point Actuator and Control Box Information Warning Possible Shock Hazard Unplug the power cord from the wall outlet before performing a...

Страница 13: ...iver Masking Tape Zip Ties 5 16 Nut Driver 1 Raise the bed so that there is about 27 inches of clearance between the floor and the mattress support platform of bed 2 Unplug the power cable from the ou...

Страница 14: ...k of use check all threaded fasteners for looseness and make sure all pins are in their normal location and fastened securely Check monthly for loose bolts nuts pins and other retaining hardware Tight...

Страница 15: ...the floor and or cause the bed to become unstable resulting in injury or death Drainage Bag Holder Warning Possible BioHazard Prior to use consult the Drain Bag manufacturer s installation and use in...

Страница 16: ...igure 36 1 The Drainage Bag Holder is designed as an attachment point for items such as urine collection bags Make sure items attached to holder are safely secured Prior to use please read important w...

Страница 17: ...be fully charged Note We recommend that users charge up the Battery immediately upon receipt 4 The Battery has a useful life of about 3 5 years depending upon usage It is recommended to discharge the...

Страница 18: ...are covered for 90 days after date of delivery Joerns Healthcare s obligation under this warranty is limited to supplying replacement parts servicing or replacing at its option any product which is fo...

Отзывы: