background image

5

Joerns® Bed Frames

EasyCare

®

 

© 2015 Joerns Healthcare • 6110158 RevC • 15-2537

Accessories And Options

76", 80" and 84" Joerns

®

 Mattresses

Joerns Bed Panels

F18 SoftCare half rail

F14SC Assist Device Set

F025 Assist Handle

F026 Assist Device

F027 Pivot Assist Handle

F028 Deluxe Assist Device

Staff Control

IV Rod

Floor Mat

Emergency Battery Backup System

Battery Backup 24V

Battery Charger 24V

• Wall Mounting Bracket for Charger

Trapeze and Trapeze Adapter

Drainage Bag Holder

Features and Specifications

Standard Features

76" or 80" Sleep Surface

Travel Range:

 

7" to 30" 

Locking System: 

Roll-in-Low Lock System

Care-Lock Roll at any Height Mobility

Pendant Jacks on both left and right side

Whisper Quiet DC Motors

Welded Tubular Frame

Grid Sleep Surface

SoftTone Frame Color

Baseboard Bumper

Wide Stable Base

Heavy-Duty Casters 

Mattress Stops

Patented Lifting Mechanism (US Patent 6473922)

EasyLift Leg Positioner

Battery Pack Cable

Electrical Specifications

Power:

 120 VAC 

Frequency: 

60 Hz

Current Rating:

 4.0 Amps (120V)

Classification: 

Class I, Type B or Class II, Type B

Mode of Operation:

 10% Maximum Duty Cycle 

2 minutes on/18 minutes off

Circuit Protection:

 LA27 Actuators: Auto Reset

Current Sensor

Environmental Conditions

Operating Conditions:

Ambient Temperature: +10°C to +40°C

Relative Humidity: 30% to 75% Non-Condensing

Atmospheric Pressure: 700 hPa to 1060 hPa

Storage and Shipping Conditions:

Ambient Temperature: -40°C to +70°C Relative

Humidity: 10% to 100% 

Atmospheric Pressure: 500 hPa to 1060 hPa

Mechanical Specifications

Description: 

Dimensions:

76"

 ........................................

80.75"L x 35.0"W x 7.0"H

80"

 ......................................

84.25"L x 35.0"W x 7.65"H

Weight

Roll-in-Low ................................................... 220 lbs.

Roll at any Height

 ......................................... 230 lbs.

Care-Fit Integrated Width.................................+40 lbs.

Maximum Back Angle .........................

65° to horizontal

Maximum Knee Angle ........................

35° to horizontal

Maximum Knee To Leg Angle ................................. 45°

Minimum Height Mattress Support Surface

Roll-in-Low ........................................................

7.00"

Roll at any Height

 ..............................................

7.65"

Maximum Height Mattress Support Surface

Roll-in-Low ......................................................

30.00"

Roll at any Height

 ............................................

30.65"

Safe Working Load .......................................... 500 lbs.

Hi-Lo Speed

Unloaded .................................................50 seconds

With 200 lbs ............................................52 seconds

Tolerance ± 3 seconds

Attention, Consult accompanying documents

I.D. Label Symbol Definition

Type B Applied Parts

Safe Working Load

Class II Equipment

 

Protective Earth (ground)

Bed Identification Numbers

When ordering parts or when contacting our 

Customer Care Department, please include the 

bed’s model, product and serial numbers, found 

on the identification labels. The identification 

labels are located on bed frame under the sleep 

deck foot section.

Содержание EasyCare

Страница 1: ...User Service Manual Joerns Bed Frames EasyCare Bed To avoid injury read user s manual before using...

Страница 2: ...rom any outlet Warning Possible Shock Hazard Injury may result from improper routing of the power cord Always follow the proper factory installed routing configuration Failure to do so could result in...

Страница 3: ...if mattress used with Mattress Side Stop does not fill the entire width between stops or which compresses to less than 1 5 inches under user s weight Warning Possible Injury or Death Floor locks incr...

Страница 4: ...Therefore compliance is the responsibility of the facility As the leading manufacturer of long term care beds and a frontrunner in addressing this critical issue Joerns Healthcare can offer you the ex...

Страница 5: ...tection LA27 Actuators Auto Reset Current Sensor Environmental Conditions Operating Conditions Ambient Temperature 10 C to 40 C Relative Humidity 30 to 75 Non Condensing Atmospheric Pressure 700 hPa t...

Страница 6: ...ain support of the bed panel Insert the safety pins on each side and latch them Figure 3 6 Press the mounting bracket against the bed frame and tighten the four panel mounting screws right and left si...

Страница 7: ...h 12 printed on it It is longer than the standard bumper 1 Remove the shorter bumper by tapping the tangs on either side of the leg with a rubber mallet 2 Place the 12 baseboard bumper 3 Tap into plac...

Страница 8: ...3 Ensure that the cable plug is fully inserted into the cable jack 4 To secure tighten the union nut on both cables Figure 14 Operation Pendant Operation Your EasyCare bed comes standard with one of t...

Страница 9: ...tion button Standard Staff Control Figure 21 2 1 4 6 5 3 7 Advanced Staff Control Figure 22 8 2 12 7 1 4 6 5 3 9 10 11 Head Operation 1 Push to raise Back 2 Push to lower Back Knee Operation 3 Push to...

Страница 10: ...Install correct baseboard bumper Bed Transport Roll in Low Lock System 1 To keep the bed from rolling when in the lowest position press down on the locking lever to lock casters Figure 23 2 The stand...

Страница 11: ...re 28 and 29 5 Release the tether to engage the detent Figure 27 6 Repeat the above steps on the remaining MSP extensions Warning Possible Injury Or Death Use a mattress that is properly sized to fit...

Страница 12: ...in intact and are securely fastened to their intended connection point Actuator and Control Box Information Warning Possible Shock Hazard Unplug the power cord from the wall outlet before performing a...

Страница 13: ...iver Masking Tape Zip Ties 5 16 Nut Driver 1 Raise the bed so that there is about 27 inches of clearance between the floor and the mattress support platform of bed 2 Unplug the power cable from the ou...

Страница 14: ...k of use check all threaded fasteners for looseness and make sure all pins are in their normal location and fastened securely Check monthly for loose bolts nuts pins and other retaining hardware Tight...

Страница 15: ...the floor and or cause the bed to become unstable resulting in injury or death Drainage Bag Holder Warning Possible BioHazard Prior to use consult the Drain Bag manufacturer s installation and use in...

Страница 16: ...igure 36 1 The Drainage Bag Holder is designed as an attachment point for items such as urine collection bags Make sure items attached to holder are safely secured Prior to use please read important w...

Страница 17: ...be fully charged Note We recommend that users charge up the Battery immediately upon receipt 4 The Battery has a useful life of about 3 5 years depending upon usage It is recommended to discharge the...

Страница 18: ...are covered for 90 days after date of delivery Joerns Healthcare s obligation under this warranty is limited to supplying replacement parts servicing or replacing at its option any product which is fo...

Отзывы: