JK2830 Operation Manual Chapter 5 Execute LCR operation and some examples
modification coefficient A
r
(refer to figure B), the current basic
measurement accuracy A is obtained by multiplying A got in step
1 with A
r
.
Figure B is the basic accuracy modification curve, where V
s
is the test signal
voltage.
[Vrms]
Ar
150m
1.2
100
50
20
10
5
2
1
3
2
1
500m
200m
100m
5m
50m
20m
10m
Drawn By :
\th2816A\th2816.ddb
产品设计图
\
赵浩华
E:\
File:
Sheet of
1-Nov-2001
Date:
B
Size
Revision
Num ber
Title
A
B
C
D
1
2
3
4
5
6
D
C
B
A
6
5
4
3
2
1
Table A Impedance rate factors: K
a
, K
b
Speed
Frequency
K
a
K
b
Medium
Slow
f
m
100Hz
−
+
+
Z
V
f
m
s
m
3
(
1 10
)(1
200
)(1
100
)
−
+
+
Z
V
f
m
s
m
9
(1 10 )(1
70
)(1
100
)
100Hz
fm
≤100kHz
−
+
Z
V
m
s
3
(
1 10
)(1
200
)
−
+
Z
V
m
s
9
(1 10 )(1
70
)
100kHz<f
m
≤200kHz
−
+
Z
V
m
s
3
(
1 10
)(2
200
)
−
+
Z
V
m
s
9
(3 10 )(1
70
)
Fast
f
m
100Hz
−
+
+
Z
V
f
m
s
m
3
(
2.5 10
)(1
400
)(1
100
)
−
+
+
Z
V
f
m
s
m
9
(2 10 )(1
100
)(1
100
)
100Hz
fm
≤100kHz
−
+
Z
V
m
s
3
(
2.5 10
)(1
400
)
−
+
Z
V
m
s
9
(2 10 )(1
100
)
100kHz<f
m
≤200kHz
−
+
Z
V
m
s
3
(
2.5 10
)(2
400
)
−
+
Z
V
m
s
9
(6 10 )(1
100
)
Where, f
m
is the test frequency and the unit is [Hz].
The unit of the tested impedance is [Ω].
The unit of the test signal voltage is [mV
rms
].
When the impedance is smaller than 500Ω, K
a
, K
b
is unavailable.
When the imp
edance is larger than 500Ω, K
a
, K
b
is unavailable.
60