JK2830 Series Operation Manual Chapter 3 Introduction to [DISP]
39
◼
OPEN OFF ON
◼
MORE 1/2
2)
Press the soft key MORE 1/2, the following softkeys will be displayed:
◼
SPOT SHORT
◼
SWEEP SHORT
◼
SHORT OFF ON
◼
MORE 2/2
3)
Connect the test fixture to the test ports. Short the test fixture by using short
plate.
4)
Press
SPOT SHORT
, JK2830 will make short correction on the current
frequency.
5)
Press the
SWEEP SHORT
soft key, JK2830 will test the short spurious
impedances (resistance and reactance) of 37 frequencies. Short full
frequency correction takes about 20 seconds and in this process, the
following soft keys will be displayed.
◼
CANCEL
This soft key can be used to cancel the current short correction operation and
reserve the formal open correction data.
Press
SHORT OFF ON
to turn ON/OFF the short correction function. If
ON,JK2830 will make short correction calculation in latter test. If OFF, no
short correction calculation will be performed in later test.
3.7
<LIST SETUP>
Press
[SET]
and then
LIST SETUP
to enter into the
<LIST SETUP>
page as
shown below.
The list sweep function of JK2830 can perform auto sweep test for the test
frequency, test level or bias voltage of 100 points. On
<LIST SETUP>
page, the
following list sweep parameters can be set.