JK2830 Series Operation Manual Chapter 3 Introduction to [DISP]
37
B
line.
11)
Prepare a standard test component.
12)
Move the cursor to
FUNC
.
13)
Set the function parameters required to be set.
14)
Move the cursor to
REF A
.
15)
Use numeric keys and unit keys to input the primary reference values of the
standard component.
16)
Move the cursor to
REF B
.
17)
Use numeric keys and unit keys to input the secondary reference value of the
standard component.
18)
Move the cursor to the corresponding
FREQ.
.
19)
Connect the standard component to the test fixture.
20)
Press
MEAS LOAD
, the instrument will execute a load correction. The real
test results of the standard component will be displayed in
LOAD A
and
LOAD B
.
21)
Move the cursor to
LOAD
.
22)
Press
ON
to perform load correction calculation at preset frequencies in latter
measurements.
3.6.4
Load correction test function
When performing load correction, the reference value of the standard component
is required to be input in advance. The test parameters of reference value should
conform with the preset load correction test function.
Load correction function adopts the transport coefficient between the real test
value of preset frequency and the standard reference value to eliminate the test
error. Load correction function is only available for calculating transport
coefficient.
Operation steps for setting load correction test function
Refer to section 3.6.3.
3.6.5
Cable length selection
The available cable length is 0m, 1m, 2m and 4m.
3.6.6
Single/ multi correction mode
Refer to the instruction of the scanting optional interface.