Chapter 5
Launching a reflectometry test and displaying results
OptiPulses option
68
User Manual
770000102/08
OptiPulses option
The OptiPulses function is delivered as a licence, available on option
with the T-BERD/MTS.
It is usable with OTDR trace exclusively and can be used before
performing an OTDR acquisition.
This option allows to merge the results of the short acquisition with those
of the standard one in the results table and summary page.
Configuring the
OTDR
acquisition with
OptiPulses
mode
Once the installation of the licence is completed (see Base-Unit manual)
and the ExpertOTDR icon is selected in the
Home
page:
1
Press
S
ETUP
hard key to display the configuration parameters
2
In the
Acquisition
parameters, select
OptiPulses
and configure
the acquisition:
No
No OTDR short acquisition with the shorted
pulsewidth is performed before the standard one
Auto
this parameter allows to perform a short
acquisition, before the standard one, with the
Short Pulse
and
Short Range
parameters
defined automatically.
Manual
this parameter allows to perform a short
acquisition, before the standard one, with the
Short Pulse
and
Short Range
parameters
defined by the user in the two following lines.
“Smart Acquisition (not available in Multimode)” on page 22
for infor-
mation on principle of the acquisition.
3
Configure the other parameters for the OTDR acquisition to be
performed
Fig. 40
OptiPulses configuration
MTS2100 4100 Module US rev08.book Page 68 Jeudi, 1. août 2013 1:43 13
Содержание 4100
Страница 2: ...MTS2100 4100 Module US rev08 book Page ii Jeudi 1 ao t 2013 1 43 13...
Страница 4: ...MTS2100 4100 Module US rev08 book Page iv Jeudi 1 ao t 2013 1 43 13...
Страница 180: ...Index 160 User Manual 770000102 08 MTS2100 4100 Module US rev08 book Page 160 Jeudi 1 ao t 2013 1 43 13...
Страница 181: ...MTS2100 4100 Module US rev08 book Page 161 Jeudi 1 ao t 2013 1 43 13...