Chapter 7
Power meter and Source options of the OTDR Modules
Activating the Source function
94
User Manual
770000102/08
wavelength, then press > to access choice of
value, then confirm this value.
Configuring the
alarm
parameters of
the power
meter
Alarm
Activation of the Alarm function : any result below
the lower threshold or above the upper threshold
will be displayed in red on the Results page.
Min and max thresholds
Choice of lower and upper thresholds for each
available wavelength, from -60 to +40 dBm
(selected with the direction keys).
Activating the Source function
The Source function is an option chosen at the time of order and incor-
porated into the OTDR module in the factory.
To activate the function:
1
Press the
H
OME
button
2
Click on the icon
Source
in the OTDR Section
The icon turns yellow
.
NOTE
To copy a Reference Level or a Attenuator Compensator on all wave-
lengths, select the reference wavelength and click on
Update for All
Wavel.
.
NOTE
To copy one value of the Lower or/and Upper threshold for all wave-
lengths, select the reference value and click on
Update for All
Wavel.
.
NOTE
A continuous push on direction keys increments the value by 10 dBm.
CAUTION
The Singlemode and Multimode Source icons cannot be selected at
the same time. When one is selected, it automatically deselects the
other one.
MTS2100 4100 Module US rev08.book Page 94 Jeudi, 1. août 2013 1:43 13
Содержание 4100
Страница 2: ...MTS2100 4100 Module US rev08 book Page ii Jeudi 1 ao t 2013 1 43 13...
Страница 4: ...MTS2100 4100 Module US rev08 book Page iv Jeudi 1 ao t 2013 1 43 13...
Страница 180: ...Index 160 User Manual 770000102 08 MTS2100 4100 Module US rev08 book Page 160 Jeudi 1 ao t 2013 1 43 13...
Страница 181: ...MTS2100 4100 Module US rev08 book Page 161 Jeudi 1 ao t 2013 1 43 13...