Chapter 8
FiberComplete Modules
Establishing References
110
User Manual
770000102/08
Connect the jumper from the module port to the non-reflective termi-
nation via a mating sleeve. If you don't have a non-reflective termi-
nation, a mandrel can be used. Press
Ok
to start referencing.
Side-by-Side
referencing
method
The side-by-side referencing is used when the two units are at the same
location and is the preferred method for better loss measurement accu-
racy.
After clicking on
Side/Side
, the wizard will guide you through three steps:
1
The side-by-side reference is used for ORL testing. Connect the
jumper from the module port to the mainframe powermeter and
press
Ok
to start referencing.
2
The zero ORL reference is necessary for ORL testing. Once the self
reference measurement has been carried out, the Zero ORL adjust-
ment can be performed
Connect the jumper from the module port to the non-reflective termi-
nation via a mating sleeve. If you don't have a non-reflective termi-
nation a mandrel can be used. Press
Ok
to start referencing.
3
For the loss reference, connect the jumper from the module port of
the primary unit, toward the module port of the secondary one via a
mating sleeve. Press
Ok
to start referencing.
Fig. 70
Zero ORL reference
Non-reflective termi-
nation (or mandrel)
OTDR port
NOTE
Non-reflective terminations are mandatory when bend insensitive
jumpers are used.
MTS2100 4100 Module US rev08.book Page 110 Jeudi, 1. août 2013 1:43 13
Содержание 4100
Страница 2: ...MTS2100 4100 Module US rev08 book Page ii Jeudi 1 ao t 2013 1 43 13...
Страница 4: ...MTS2100 4100 Module US rev08 book Page iv Jeudi 1 ao t 2013 1 43 13...
Страница 180: ...Index 160 User Manual 770000102 08 MTS2100 4100 Module US rev08 book Page 160 Jeudi 1 ao t 2013 1 43 13...
Страница 181: ...MTS2100 4100 Module US rev08 book Page 161 Jeudi 1 ao t 2013 1 43 13...