Chapter 5
Launching a reflectometry test and displaying results
Performing OTDR acquisitions
42
User Manual
770000102/08
Performing an
acquisition
from Results
page
Once the results page is displayed, you can perform a new acquisition
modifying the main acquisition parameters.
1
On Results page, press
Quick Setup
softkey
In OTDR Expert mode, the acquisition parameters that can be modi-
fied displays under the results trace.
In SmartOTDR™ mode, the Setup screen displays (see
).
2
Modify the acquisition parameters wished in the displayed menu:
–
In OTDR Expert mode: Laser / Acquisition / Range / Pulse /
Time
and
.
–
In SmartOTDR™ mode: Laser / Fiber Number / Distance unit
.
3
Once configured, launch the new OTDR test pressing the
S
TART
/
S
TOP
hard key.
Press again
Mini Setup
menu key to hide the menu under the trace.
Before launching a new OTDR acquisition, make sure the
trace(s) displayed have been previously saved if necessary, as
the new acquisition will automatically delete the displayed
results.
Fig. 20
Results page and Quick Setup menu (OTDR Expert)