background image

11

HAND CHECK TEST

Test the mattress’s function with a simple hand check.

To ensure that the mattress is providing the right therapeutic function, a simple hand check is 
performed. The test should be performed each time a mattress is taken into operation in order to 
detect any problems with the mattress’s air pressure. The test ensures that the patient does not 
“bottom out”.

How the hand check is performed.

The hand check should always be performed when the patient is positioned on the mattress.  
It is important to ask the patient not to try and lift their body during the test and to stay lying as 
usual. It is not uncommon for patients to lift their body in the belief that it is easier for the staff, 
but it may give an incorrect test result.

•  Fold away the sheets and unzip the mattress cover to access the cells.

•  Put one hand between two cells with the patient lying on the mattress. 

(See illustration)

.

Depending on the size of the hand, it may sometimes be sufficient to use 2-3 finger height.

•  Ensure that when the hand is in position, it fits between the bed base and the patient.

The important conclusion of the check is that there is space between the patient and the 
bed base.

If the hand check shows that the patient is “bottoming out”.

If the patient is “bottoming out” and no alarm has triggered on the pump, the comfort level

 

should be raised 

(position B on page 14)

. If it is already set to maximum and the patient does 

not weigh more than the recommended maximum weight, then the mattress and pump need 
troubleshooting by authorised personnel (see page 12).

Содержание OptiCell 4 PRO

Страница 1: ...1 USER GUIDE OptiCell 4 PRO Alternating pressure air mattress...

Страница 2: ...roduct 8 Receiving and unpacking 8 Storage 8 Automatic adjustment and alternating pressure 9 Static mode 9 Low pressure alarm 9 Air hose connection 10 Hand Check test 11 Inspection and troubleshooting...

Страница 3: ...e CE marking signifies that OptiCell AB has the manufacturer s responsibility to ensure that this product meets the essential requirements of relevant European legislation This includes the provision...

Страница 4: ...ment of the intended usage environment and then use the right product Please contact us if you have any questions In order to avoid damage the mattress should not come into contact with heat sources o...

Страница 5: ...use maintenance and improved safety It is very important to read this guide carefully before using the product Care personnel must be given access to this user guide and also be informed of the risks...

Страница 6: ...the bi elastic surface layer reduces the risk of shear and other complications for the patient Fluid protection flap and waterproof zip on four sides with welded seams Anti slip underside which grips...

Страница 7: ...ell 4 PRO air mattress is connected to the pump by three air hoses The red air hose positioned in the middle is used to fill the head cells situated under the patient s head It also fills the lower ch...

Страница 8: ...n mind that a tall patient can weigh the same as a short patient but may need a different pressure This must always be carefully observed for each patient The biggest risk for the patient is to bottom...

Страница 9: ...not want to be disturbed by an alternating pressure Press the Care Static button position E on page 14 to select either Care mode or Static mode Care also called Auto firm mode provides maximum pressu...

Страница 10: ...connector if the air hose is disconnected from the pump To minimise the risk of air leakage during hose connection check That the two hooks are not damaged and their clips are intact That the black o...

Страница 11: ...t is easier for the staff but it may give an incorrect test result Fold away the sheets and unzip the mattress cover to access the cells Put one hand between two cells with the patient lying on the ma...

Страница 12: ...r hoses for signs of damage or abnormal wear Check the air hose connector see page 10 INSPECTION OFTHE OPTITEX COVER Regularly check the cover for visible signs of damage or wear such as tears cracks...

Страница 13: ...d at a minimum of 70 C for at least 10 minutes with unscented detergent according to standard BS EN 14065 2016 Textiles Laundry Processed Textiles Biocontamination Control System The OptiTex cover can...

Страница 14: ...ng on the mattress is affected by the comfort mode chosen Press this button to cycle through modes ALARM If the alarm sounds press this button to mute and reset The adjacent Leaking lamp will light up...

Страница 15: ...out condensation Storage 10 90 without condensation Safety standard CE marked CARE STATIC Press this button to select Care auto firm mode or Static mode When the adjacent Care lamp is lit the pump pro...

Страница 16: ...posed of at a recycling station and sorted as Combustible waste For more information contact J rven HealthCare J rven Health Care takes environmental responsibility seriously We strive to always use m...

Страница 17: ...COMPLAINTS Any complaints regarding the product must be notified in writing to J rven Health Care The return of products must be done in accordance with our instructions Returning a product before a c...

Страница 18: ...essment is made to identify the specific needs of the patient regarding pressure relief and must be conducted by staff with adequate knowledge and education Based on the patient s assessed status the...

Страница 19: ...19 YOUR NOTES If you require any further information or assistance please get in touch with us We will be happy to help you...

Страница 20: ...e OPTICELL 4 PRO user guide All trademarks are the property of their respective owners J rven AB reserves the right to update this guide if necessary and or on finding any spelling or factual errors A...

Отзывы: