background image

9

The mattress can be used immediately after delivery and unpacking. Inflating the mattress 
takes about 15 minutes and should be done before the patient is placed on the mattress.

AUTOMATIC ADJUSTMENT AND ALTERNATING PRESSURE

The pump will automatically adjust the air pressure based on the patient’s weight, length, location, 
movements and position on the mattress. This feature makes it easy to start using the mattress.

The comfort level can be adjusted by raising or lowering the general air pressure. This is done 
by pressing the 

Comfort

” 

button 

(position B on page 14)

. Ensure that the patient does not 

“bottom out”. See Hand Check test on page 11.

The time interval for alternating pressure is selected by pressing the 

”Cycle Time”

 button 

(position D on page 14)

. The patient’s experience of the mattress and their pressure ulcer needs 

are the basis for choosing the time interval.

STATIC MODE

In some cases, it is necessary to switch the mattress’s air pressure from alternating to static.  
It can be for instance, during nursing care procedures or when the patient does not want to 
be disturbed by an alternating pressure. Press the 

“Care/Static” 

button 

(position E on page 14)

 

to select either 

“Care” 

mode or 

“Static”

 mode:

• 

“Care” 

(also called Auto-firm mode) provides maximum pressure to facilitate nursing care.

• 

“Static” 

adjusts the air pressure after setting the comfort mode.

The correct therapeutic effect is achieved with alternating air pressure in Dynamic 
mode. Static air pressure (constant low pressure) should only be used when necessary.

LOW PRESSURE ALARM

If the mattress air pressure falls too low, the 

“Leaking” 

lamp lights up and an audio alarm starts.

To respond, start by pressing the 

“Alarm” 

button 

(position C on page 14) 

to silence the alarm 

and then check the following points to find the cause.

•  Check that the air hose is connected correctly to the pump.

•  Check that no air is leaking from the valves connecting the air cells and air hose.

•  Check that the CPR valve is properly attached.

•  Check all air cells for leakage. This is easiest to do when the pump is set to 

“Care” 

(Auto-

firm mode) and all cells are properly connected.

If the alarm still persists and the 

“Leaking”

 lamp does not switch off, the mattress should be 

replaced. Contact Järven Health Care for consultation and advice – information can be found 
on the back cover of this manual.

Содержание OptiCell 4 PRO

Страница 1: ...1 USER GUIDE OptiCell 4 PRO Alternating pressure air mattress...

Страница 2: ...roduct 8 Receiving and unpacking 8 Storage 8 Automatic adjustment and alternating pressure 9 Static mode 9 Low pressure alarm 9 Air hose connection 10 Hand Check test 11 Inspection and troubleshooting...

Страница 3: ...e CE marking signifies that OptiCell AB has the manufacturer s responsibility to ensure that this product meets the essential requirements of relevant European legislation This includes the provision...

Страница 4: ...ment of the intended usage environment and then use the right product Please contact us if you have any questions In order to avoid damage the mattress should not come into contact with heat sources o...

Страница 5: ...use maintenance and improved safety It is very important to read this guide carefully before using the product Care personnel must be given access to this user guide and also be informed of the risks...

Страница 6: ...the bi elastic surface layer reduces the risk of shear and other complications for the patient Fluid protection flap and waterproof zip on four sides with welded seams Anti slip underside which grips...

Страница 7: ...ell 4 PRO air mattress is connected to the pump by three air hoses The red air hose positioned in the middle is used to fill the head cells situated under the patient s head It also fills the lower ch...

Страница 8: ...n mind that a tall patient can weigh the same as a short patient but may need a different pressure This must always be carefully observed for each patient The biggest risk for the patient is to bottom...

Страница 9: ...not want to be disturbed by an alternating pressure Press the Care Static button position E on page 14 to select either Care mode or Static mode Care also called Auto firm mode provides maximum pressu...

Страница 10: ...connector if the air hose is disconnected from the pump To minimise the risk of air leakage during hose connection check That the two hooks are not damaged and their clips are intact That the black o...

Страница 11: ...t is easier for the staff but it may give an incorrect test result Fold away the sheets and unzip the mattress cover to access the cells Put one hand between two cells with the patient lying on the ma...

Страница 12: ...r hoses for signs of damage or abnormal wear Check the air hose connector see page 10 INSPECTION OFTHE OPTITEX COVER Regularly check the cover for visible signs of damage or wear such as tears cracks...

Страница 13: ...d at a minimum of 70 C for at least 10 minutes with unscented detergent according to standard BS EN 14065 2016 Textiles Laundry Processed Textiles Biocontamination Control System The OptiTex cover can...

Страница 14: ...ng on the mattress is affected by the comfort mode chosen Press this button to cycle through modes ALARM If the alarm sounds press this button to mute and reset The adjacent Leaking lamp will light up...

Страница 15: ...out condensation Storage 10 90 without condensation Safety standard CE marked CARE STATIC Press this button to select Care auto firm mode or Static mode When the adjacent Care lamp is lit the pump pro...

Страница 16: ...posed of at a recycling station and sorted as Combustible waste For more information contact J rven HealthCare J rven Health Care takes environmental responsibility seriously We strive to always use m...

Страница 17: ...COMPLAINTS Any complaints regarding the product must be notified in writing to J rven Health Care The return of products must be done in accordance with our instructions Returning a product before a c...

Страница 18: ...essment is made to identify the specific needs of the patient regarding pressure relief and must be conducted by staff with adequate knowledge and education Based on the patient s assessed status the...

Страница 19: ...19 YOUR NOTES If you require any further information or assistance please get in touch with us We will be happy to help you...

Страница 20: ...e OPTICELL 4 PRO user guide All trademarks are the property of their respective owners J rven AB reserves the right to update this guide if necessary and or on finding any spelling or factual errors A...

Отзывы: