Step 3:
Adjusting of sensor
direction
The Open Channel transducer angle is fixed to 60°. The angle is measured from the
CENTER of the sensor to the mounting wall.
In order to receive a good signal for best accuracy, all transducers must face diagonally.
For trapezoidal or circular channels, rotate the sensors on both walls so that they face
each other.
Step 4:
Install Flowmeter
and Connect
Cables are integrated with transducers, so connect the other side of cable to the flow
computer.