2
Hexagon wrench
Bolt cover
Remove the bolt cover by a 3mm hexagon wrench (TC/TE types).
Use a 13mm or a 8mm spanner to remove the pump head
fi
xing bolts (eight M8 bolts for the C150/ six M5 bolts for
C060) and detach the pump head with a reinforcing plate.
Extend the pump shaft by keypad operation.
Select "Diaphragm Pos." through the menu screen and
push the Enter key. See page 47 for detail.
Select "MAX OUT Pos." by the right and left keys and push
the Enter key.
The diaphragm extends to the maximum and "MAX OUT
Pos." blinks on the screen.
Rotate and remove the diaphragm and the retainer.
NOTE
Be careful not to be scratched by the pump head holders.
Wear part replacement
5
4
Reinforcing plate (TC/TE types)
Pump head
3
Pump shaft
Diaphragm
Retainer
Pump head holder