Ivoclar Vivadent Programat EP 5000 G2 Скачать руководство пользователя страница 36

36

8.2   Technical malfunctions

These malfunctions may occur without an error message being displayed.
* If there are any questions, please contact the Ivoclar Vivadent After Sales Service. 

8.3  Repair

  Repairs may only be carried out by a certified Ivoclar 

Vivadent Service Center. Please refer to the addresses on 
the last page of these Operating Instructions.

If repairs during the warranty period are not carried out by a certified 
Ivoclar Vivadent Service Center, the warranty will be voided  
immediately. Please also refer to the corresponding warranty  
regulations.

8.4  Load factory settings 

If you want to reset the furnace to its original settings, choose  

Selection - Settings - Extended settings - Factory settings

. All 

programs, OSD settings, melodies, navigation settings and volume 
control settings will be reset to the factory settings.

Description

Double-check

Action

 

Vacuum is not released or only very slowly

Is the vacuum released within approximately 30 
seconds?

Wait until the vacuum is released, remove object. 
Switch the furnace on and off again * 

Indication on display incomplete

Activate the display test program 

*

Writing in the display is very hard to read

Is the contrast properly set?

Adjust contrast

Display not illuminated

Is the furnace properly connected according to the 
Operating Instructions and switched on?

Correctly connect the furnace and switch it on

Buzzer does not sound

Is the buzzer switched off (Tune 0)?

Select tune 1–5

Furnace head does not open

Was the furnace head moved manually?

Open the furnace head only by using the  
corresponding keys. Switch the furnace on and off 
again.

Has the vacuum already been released?

Is the program still running? Wait until the program 
is complete. Switch furnace off and on again. *

Vacuum pump does not start working

Is the vacuum pump fuse defective?

Check fuse and replace if necessary

Was the maximum power consumption exceeded?

Use only the vacuum pump recommended by 
 Ivoclar Vivadent

Is the vacuum pump plug correctly connected?

Correctly connect the vacuum pump to the furnace 
base.

Final vacuum is not reached

Is the vacuum hose OK?

Check vacuum hose and hose connection.

Is the pump output OK?

Start the vacuum test program

Humidity/condensation in the vacuum hose?

Start dehumidification program

Incorrect or illogical temperature indication

Is the thermocouple bent or fractured?

*

Is the thermocouple correctly connected?

Correctly connect thermocouple

Is the thermocouple plug defective?

*

Hairline cracks in the heating muffle

Are the cracks very small and insignificant (hairline 
cracks)?

Small cracks in the muffle are normal and do not 
negatively influence the function of the furnace.

Are the cracks large or have parts of the heating 
muffle broken off?

*

Cracks in the insulation

Are the cracks very small and insignificant (hairline 
cracks)?

Small cracks in the insulation do not negatively 
influence the furnace

Are the cracks large or have parts of the insulation 
broken off?

*

Cracks in the quartz glass / heating element

Are there cracks in the quartz glass or is the quartz 
glass sheathing the heating wire broken?

Switch off the furnace. *

Содержание Programat EP 5000 G2

Страница 1: ...Operating Instructions Programat EP 5000 G2 Valid as of Software Version 6 0...

Страница 2: ...2...

Страница 3: ...tructure 5 4 Adjustable parameters and possible value ranges 5 5 Settings and information 5 6 Explanation of the symbols on the display 5 7 Explanation of the speaker signals 5 8 OSD Optical Status Di...

Страница 4: ...screw for hood 38 Air vents furnace head 39 Air vents rear panel 40 Warnings 41 Furnace head mounting mark 42 Furnace base mounting mark 43 Furnace head mounting 44 Quartz glass tube 46 Vacuum hose 4...

Страница 5: ...5 40 38 56 34 6 10 35 51 7 1 2 3 9 5 8 55 23 33 36 58 20 63 52...

Страница 6: ...6 49 66 26 30 60 67 29 27 28 11 32 43 22 61 63 59 22 64 62 65 11 36 37 39 12 13 53 57 19 12 11 13 15 19 53 57 21 17 18...

Страница 7: ...ead 92 Numeric keys 93 Firing Pressing 94 Home key 95 Help key 96 Power Saving Key 8 59 4 44 3 50 54 41 42 100 Programat firing tray 101 Metal pin A 102 Metal pin B 103 Metal pin C 115 Cooling grid co...

Страница 8: ...tion Burn hazard Risk of crushing The Operating Instructions must be read 1 3 Notes regarding the Operating Instructions Furnace concerned Programat EP 5000 G2 Target group Dental technologists These...

Страница 9: ...structions for Use must be observed The furnace must be operated under the indicated environmental and operating conditions Chapter 9 The Programat EP 5000 G2 must be properly maintained 2 1 1 Risks a...

Страница 10: ...a burn hazard 2 1 12 Contraindication Do not insert any foreign objects into the air vents There is a risk of electrical shock 2 1 13 Contraindication This product contains ceramic fibres and may rele...

Страница 11: ...r must be disconnected to avoid accidental operation Safe operation is no longer possible if the furnace is visibly damaged the furnace does not work the furnace has been stored under unfavourable con...

Страница 12: ...ed by a press drive The firing pressing programs are controlled with the corresponding electronic controls and software Moreover the set and actual temperatures are continuously compared 3 4 Accessori...

Страница 13: ...ents Simply use the two corresponding inserts Fold the side flaps 2 and combine the two packaging parts by means of the transportation flaps The packaging may be disposed with the regular household re...

Страница 14: ...user Lift the furnace head with both hands see picture and carefully position it on the furnace head mounting 43 Secure the cooling tray 34 with the two screws 35 including the silicone washer 47 Plac...

Страница 15: ...aligned with the furnace base mounting mark 42 Make sure that the firing plate holder 48 is not damaged by mounting the furnace head Step 4 Connections Connect the cables of the furnace head with the...

Страница 16: ...blishing additional connections Power connection Please make sure that the voltage indicated on the rating plate complies with the local power supply Connect the power cord with the power socket 17 of...

Страница 17: ...r socket 17 1 Loosen and remove the knurled screw 37 of the hood 36 2 Remove the hood 36 3 Disconnect the thermocouple plug 26 4 Disconnect the heater plug 28 5 Press the leaf spring 32 with a finger...

Страница 18: ...n will be displayed The desired language is set by means of the respective touch button After that the next basic setting screen temperature mode appears These settings are then saved and will no long...

Страница 19: ...Power saving key 89 Power saving function activated only possible with the furnace head closed and the furnace on idle The display shows the power saving icon Pressing any key ends the power saving fu...

Страница 20: ...SP is only active if no active predrying has been selected or if the temperature in the firing chamber is too high when the program is started In addition TSP works only if the stand by temperature of...

Страница 21: ...touch button Laboratory name the keypad is displayed and the name of the laboratory can be entered With Active table and PC the protocol can be sent to the connected PC software instead of being prin...

Страница 22: ...ible with the furnace head closed and the furnace on idle The display shows the power saving icon Pressing any key ends the power saving function The energy saving mode Power Saving Technology is avai...

Страница 23: ...memory stick Select Selection Program manager Copy programs and make a choice using the keys All programs Program copy a single program Group copy a single group When a USB memory stick is operational...

Страница 24: ...e with the ESC key the signal transmitter is also switched off Symbol Name Meaning Symbol Pre vacuum Vacuum generation starts before the heating begins One stage program Abstract firing curve of a one...

Страница 25: ...ettings OSD system 5 8 3 Deactivation In order to deactivate the OSD the brightness has to be set to 0 Colour Activity white blinking The furnace is in the switch on mode self test active white Stand...

Страница 26: ...program has been set to one stage 6 2 2 Operating indicator Firing curve If the program is started with the START key the firing curve display with the vacuum status is shown The following informatio...

Страница 27: ...program group of your choice Now select the desired program e g 1st opaquer Step 2 Now open the furnace head by pressing the Open furnace head key 90 and place the firing tray with the object to be f...

Страница 28: ...etween the Stand by and Parameters screen The shown parameters are displayed for information purposes and cannot be edited Operation screen during a press program in the IPF mode 6 3 4 CDS Crack Detec...

Страница 29: ...TOP key once to pause a running program The green LED in the START key blinks Press the STOP key twice to completely stop the program or press START to continue 6 4 5 Changing the parameters while the...

Страница 30: ...t program function has been activated the furnace head remains open after the firing process the heater is switched off and the green START LED flashes No melody is played if the temperature drops bel...

Страница 31: ...e also observe the shorter service and maintenance intervals What Part When Check all plug in connections for correct fit Var external connections weekly Check if the furnace head opens smoothly and w...

Страница 32: ...Two different chequer board patterns are alternately shown on the entire display This allows the visual check of each individual pix el The display test can be ended by pressing ESC 7 4 Temperature ca...

Страница 33: ...ng movements from the split taper socket 64 with one hand and pull down the press plunger in the firing chamber with the other hand Contraindication Do not touch the thermocouple when replacing the pr...

Страница 34: ...ot to start a firing program 27 no Furnace head cannot be initi alized The furnace head cannot be moved to the final position It might be blocked by an external mechanical source If not contact your S...

Страница 35: ...e memory stick in pro gress memory stick no longer present A program in progress started from the USB memory stick was interrupted by a power failure The program could not be continued since the USB m...

Страница 36: ...head does not open Was the furnace head moved manually Open the furnace head only by using the corresponding keys Switch the furnace on and off again Has the vacuum already been released Is the progr...

Страница 37: ...ions of the closed furnace Depth 470 mm width 305 mm 400 mm with cooling tray Height 565 mm Usable size of the firing chamber Diameter 80 mm Height 48 mm Max firing temperature 1200 C Weight Furnace b...

Страница 38: ...nt Service Center Important information The current program table is also available at www ivoclarvivadent com downloadcenter The program tables can be downloaded from the Internet as PDF files Please...

Страница 39: ...39 10 2 Menu structure 10 2 1 Possibilities of the program selection Firing programs Press programs...

Страница 40: ...40 10 2 3 Overview of program groups 10 2 2 Adjustments Information...

Страница 41: ......

Страница 42: ......

Страница 43: ......

Страница 44: ...3 Piso 14 Col Napoles 03810 M xico D F M xico Tel 52 55 50 62 10 00 Fax 52 55 50 62 10 29 www ivoclarvivadent com mx Ivoclar Vivadent Ltd 12 Omega St Albany PO Box 5243 Wellesley St Auckland New Zeala...

Отзывы: