35
Please contact the Ivoclar Vivadent After Sales Service, if one of the
following error messages is being displayed:
25, 29
43, 44, 45, 46, 47, 48
54, 56
103, 107
143, 144, 145, 146, 147, 148
700, 701, 703, 704, 705, 706, 707
1010, 1011, 1012, 1013, 1014, 1015, 1016, 1017, 1018, 1019
1024, 1025, 1026, 1028
1143, 1144, 1145, 1146, 1147, 1148
1202, 1203, 1204, 1205, 1206, 1207
1300, 1301
1400, 1401, 1402
1500, 1530
*
Furnace head opens when this error occurs.
** A program in progress is stopped.
*** The error cannot be acknowledged; the programs cannot be started.
Error /
Hint
Nr.
Conti-
nuation
possible
Error
Error Message Text
520
Error muffle crack
CDS Crack Detection System has been activated. The program has been aborted and the press
plunger has been moved backwards. CDS could probably save your restorations from muffle cracks.
Please check your restorations before you continue your working progress.
521
Error muffle crack
CDS Crack Detection System has been activated. The program has been aborted and the press
plunger has been moved backwards. CDS could probably save your restorations from muffle cracks.
Please check your restorations before you continue your working progress.
522
Error muffle crack
CDS Crack Detection System has been activated. The program has been aborted and the press
plunger has been moved backwards. CDS could probably save your restorations from muffle cracks.
Please check your restorations before you continue your working progress.
526
T is > B + 200°C at the start of
a IPF press progr.
Firing chamber is too hot for the start of a IPF press program.
527
IPF press program parameters
not plausible
The set IPF press program parameters are not plausible.
528
T is < 350°C at the start of a IPF
press program
The furnace is too deeply cooled down for the start of a IPF press program.
529
IPF press program Error
temperature
The temperature during the IPF press program does not comply with the specifications. The
program may be continued nonetheless.
530
Error during logging of press
program data
An error has occurred during logging of press program data. The storage medium might be full.
700
Supply voltage outside the
acceptable range
The supply voltage is outside the acceptable range. Check the supply voltage.
701
***
no
Start-up aborted due to an error The self-test of the furnace was interrupted by an error. It is not possible to work with the furnace!
Switch the furnace off and on again, once the error has been rectified.
702
Brief power failure during of a
program in progress
A program in progress was interrupted by a brief power failure. The program is continued!
703
Power failure during a program
from the memory stick in pro-
gress memory stick no longer
present
A program in progress (started from the USB memory stick) was interrupted by a power failure.
The program could not be continued, since the USB memory stick is no longer present!
704
Power failure during an over-
night program in progress
An overnight program in process was interrupted by a power failure. The program is continued!
707
Incorrect supply voltage
The furnace is operated with the incorrect supply voltage. Make sure that the furnace is operated
with the supply voltage indicated on the rating plate.
800
Final vacuum value not reached
The required final vacuum value cannot be reached. Check the vacuum pump.
801
Vacuum drop
An unacceptable vacuum drop has occurred.
802
The vacuum does not increase
(self-test)
No vacuum increase could be measured. Check the following points: Is the firing chamber tight
(no contamination on the sealing surfaces)? Is the vacuum hose connected? Is the vacuum pump
connected? Is the fuse F1 o.k.?
1302
**
ATK2 calibration:
Pre-heating to 962 °C
Error during calibration. Sample may not be correctly inserted. Try again with a new sample and
make sure the sample makes ample contact.
1310
Calibration reminder
Some time has passed since the last calibration procedure. Calibrate the furance soon.
1311
Calibration reminder – press
cycles
Some press cycles has been done since the last calibration procedure. Calibrate the furance soon.
1312
Dehumidification reminder
Some time has passed since the last dehumidification. Conduct a dehumidification in the near
future.
1510
Temperature > VT at the start
of a firing program
The temperature in the firing chamber is higher than the pre-drying temperature. Press START to
continue the program despite the error message.
1520
Software update: Invalid file
The file selected for the software update is invalid. The update cannot be conducted.
1521
Software update: Invalid version
The software to be installed is older or the same as the one installed on the furnace. The update
cannot be conducted
1522
Software update: Error during
update
An error has occurred during the software update. Do NOT switch off the furnace and try again. If
the error reoccurs, try conducting the update via the USB interface.
1550
Change operation mode
The operation mode has been changed! Make sure that the furnace runs at the new stand-by
temperature long enough before you start a program.
1700
Service note
Two years have passed or the heating muffle has been in use for more than 1200 firing hours since
the last inspection of the furnace. Ivoclar Vivadent therefore recommends that your furnace should
be inspected. Please refer to the Equipment Service Passport or the Operating Instructions for
further information. The interval until the next appearance of the note can be set in the Settings.
Содержание Programat EP 5000 G2
Страница 1: ...Operating Instructions Programat EP 5000 G2 Valid as of Software Version 6 0...
Страница 2: ...2...
Страница 5: ...5 40 38 56 34 6 10 35 51 7 1 2 3 9 5 8 55 23 33 36 58 20 63 52...
Страница 39: ...39 10 2 Menu structure 10 2 1 Possibilities of the program selection Firing programs Press programs...
Страница 40: ...40 10 2 3 Overview of program groups 10 2 2 Adjustments Information...
Страница 41: ......
Страница 42: ......
Страница 43: ......