34
8. What if ...
This chapter will help you to recognize malfunctions and take
appropriate measures or, if possible and acceptable, to perform
some simple repairs.
8.1 Error messages
The furnace continuously checks
all functions during operation. If an
error is detected, the respective
error message is displayed.
The following error messages may be displayed. If there are any questions, please contact the Ivoclar Vivadent After Sales Service.
Error /
Hint
No.
Conti-
nuation
possible
Error
Error Message Text
2
T < B
Enter a logical value for T
8
L > T
Enter a logical value for long-term cooling L
9
V2x <= V1x
Enter a logical value for the vacuum-on temperature Vx1 or the vacuum-off temperature Vx2
10
V2x > Tx + 1°C
Change either the vacuum values or the holding time T
11
Incorrect values for V1x, V2x
Enter a logical value for V1x, V2x
13
*,**
Current temperature after Start
> Tx + 80°C
Excess temperature! Program aborted, furnace head opens to allow the furnace to cool down.
16
T2 < T1
Enter a lower value for T1 or a higher value for T2
17
Power failure > 10 s during a
firing program in progress
A firing program in progress was interrupted for more than 10 s. The program cannot be continued!
18
T1 > V12
Enter a lower value for T1 or a higher value for V12
19
vV set, but V2 is missing or
invalid
Pre-vacuum activated! V2 must be higher than B.
20
**
no
Error in the heating system
Check the heater fuse. If the fuse is i.o. contact your Service Center.
23
Heating muffle very old
The heating muffle is very old. It is recommended to replace it. After the error message has been
acknowledged, a firing program may still be started
24
Heating muffle defective
The condition of the muffle is so poor that is has to be replaced immediately.
26
T is > B + 160 °C at the start of
a firing program
Firing chamber too hot to start a firing program.
27
**,***
no
Furnace head cannot be initi-
alized
The furnace head cannot be moved to the final position. It might be blocked by an external
mechanical source. If not contact your Service Center.
28
**
The furnace head does not
reach the target position
The furnace head does not open/close correctly. The furnace head was manually moved or is
obstructed. The furnace head must only be moved using the keys intended for this purpose!
32
**
no
The vacuum is not released
The vacuum cannot be released. The vacuum valve might be dirty or stuck. Contact your Service
Center.
33
Necessary vacuum (xxxmbar) is
not reached within 1 min.
The vacuum cannot be established. Check the seal of the firing chamber, vacuum hose, vacuum
pump, pump fuse.
103
Program start blocked
Starting a program is not possible due to a technical malfunction.
106
Timer activated – unit into
stand-by mode
The program cannot be started, because the unit was put into stand-by mode by the timer (heating
turned mode off). Deactivate the timer or extend the time frame to be able to execute programs
107
Incorrect time settings
(date / time)
The clock settings are incorrect. Please set a correct date and a correct time!
108
Print error
An error occurred during printing.
110
HV > H (H2)
Enter a lower value for HV or a higher value for H (H2).
111
Maximum number of firing pro-
gram protocol entries reached
The maximum number of firing program protocol entries has been reached. For the next protocol,
an existing entry will be deleted / overwritten.
120
Share of the holding time with
vacuum“ is activated, but Vx2
does not correspond to Tx or
Tx+1
Activate the vacuum during the holding time Tx or deactivate HV.
500
Error pressing time
Max. pressing time exceeded
504
Error press position
Max. position exceeded
505
Error press force
Max. press force exceeded
513
Error press drive initalized
Press drive is not initialized. Please switch the furnace off and on again.
514
Error press drive
Technical error in press drive
Содержание Programat EP 5000 G2
Страница 1: ...Operating Instructions Programat EP 5000 G2 Valid as of Software Version 6 0...
Страница 2: ...2...
Страница 5: ...5 40 38 56 34 6 10 35 51 7 1 2 3 9 5 8 55 23 33 36 58 20 63 52...
Страница 39: ...39 10 2 Menu structure 10 2 1 Possibilities of the program selection Firing programs Press programs...
Страница 40: ...40 10 2 3 Overview of program groups 10 2 2 Adjustments Information...
Страница 41: ......
Страница 42: ......
Страница 43: ......