6.4.16 USB printer 

Each USB-PCL printer can be used to printout the protocol. If a USB-
PCL printer is connected to the furnace, the necessary software  
driver is loaded. After that, the USB printer is immediately ready for 
use. 
–  The status of the USB-PCL printer is shown in the „Information“ 

menu item

–  The desired protocol to be printed can be selected in the  

protocol table.

6.4.17 USB memory stick 

Most USB memory sticks may be 
used to store programs. For that 
purpose, the USB memory stick 
must be prepared as an external 
 program memory. 
Once a prepared USB memory   
stick is ready for use, the program 
selection screen will show another 
touch button for the selection of 
the UBS memory stick. 
–  The active status of the USB memory stick is shown in the   

“Information“ menu item (page 3/3).

–  In the Program Manager, the USB memory stick may now be  

selected as either the source and/or the target.

In this way, a backup copy of all the programs can now be made at 
any time.  

6.4.18 Timer 

The timer can be used to put the 
furnace in an energy saving mode 
for a defined period of time.  
During this time, the heating of  
the furnace is deactivated, so that 
the power consumption is only 
minimal. The furnace itself,  
however, remains switched on. 
Avoid activating the timer when  
the furnace head is open. 

30

6.4.12 Pre-vacuum

If a firing program with pre- 
vacuum is conducted, the vacuum 
pump is switched on at the end of 
the closing time (as soon as the 
furnace head is closed). Once the 
pre-vacuum time has elapsed, the 
heating phase begins. Upon the 
start of a program with an  
individually activated pre-vacuum 
(value between 1:00 and 5:00), 

the V1 value is ignored. The vacuum is maintained until V2 is  
reached. V2 must be higher than the stand-by temperature B.

6.4.13 Overnight program 

The overnight program function can be activated for the next  
program sequence in the parameter list (touch button with the  
„crescent moon“ symbol). 
If the overnight program function has been activated, the furnace 
head remains open after the firing process, the heater is switched 
off, and the green START LED flashes. 
No melody is played if the temperature drops below a certain level. 
Once the temperature has dropped below 100 °C / 212 °F, the  
furnace head closes, the heater remains switched off, and the  
furnace cools to room temperature. 
Advantages of the overnight program function:
In case of a power failure, the overnight program is always  
continued. The program resumes at the point at which the power 
failure has occurred. 

6.4.14 ”Quick cooling“ 

If the „Open furnace head“ key is pressed again with the furnace 
head already completely open, the „Quick cooling“ function is  
started. This means that the vacuum pump is switched on for  
5 minutes. This function can be stopped early by pressing STOP, 
„Close furnace head“, or START.

6.4.15 Software Update

The user will be able to conduct a software update by CD, PC, and 
download cable. For that purpose, the software download mode of 
the furnace is activated by pressing two special keys simultaneously 
while the power supply is switched on. For further details, please 
refer to the download center (www.ivoclarvivadent.com/  
downloadcenter).

A further method for updating the software is to use the USB 
 memory stick. This item is available in the Settings menu on page 2.
To update the software, a USB memory stick is required with a 
 current software image in the form of a *.gz file. The software 
 version on the USB memory stick must be newer than the one on 
the furnace (see Selection – Information).
Step 1:   Connect the USB memory stick with the current software  
 

version to the furnace

Step 2:   Select the menu item “Software Update”
Step 3:   Select the software image file (the file name has to start  
 

with EP 5000/G2_)

Step 4:   Start the software update with the “Open” key
Step 5:   Wait until the furnace indicates a successful update  
 

process

Step 6:   Re-start the furnace

  Important

 

 Please note that the individual programs will be over- 
written in the software update process. Therefore, we 
 recommend a data back-up (e.g. on a USB stick or with 
software PrograBase 2).

Содержание Programat EP 5000 G2

Страница 1: ...Operating Instructions Programat EP 5000 G2 Valid as of Software Version 6 0...

Страница 2: ...2...

Страница 3: ...tructure 5 4 Adjustable parameters and possible value ranges 5 5 Settings and information 5 6 Explanation of the symbols on the display 5 7 Explanation of the speaker signals 5 8 OSD Optical Status Di...

Страница 4: ...screw for hood 38 Air vents furnace head 39 Air vents rear panel 40 Warnings 41 Furnace head mounting mark 42 Furnace base mounting mark 43 Furnace head mounting 44 Quartz glass tube 46 Vacuum hose 4...

Страница 5: ...5 40 38 56 34 6 10 35 51 7 1 2 3 9 5 8 55 23 33 36 58 20 63 52...

Страница 6: ...6 49 66 26 30 60 67 29 27 28 11 32 43 22 61 63 59 22 64 62 65 11 36 37 39 12 13 53 57 19 12 11 13 15 19 53 57 21 17 18...

Страница 7: ...ead 92 Numeric keys 93 Firing Pressing 94 Home key 95 Help key 96 Power Saving Key 8 59 4 44 3 50 54 41 42 100 Programat firing tray 101 Metal pin A 102 Metal pin B 103 Metal pin C 115 Cooling grid co...

Страница 8: ...tion Burn hazard Risk of crushing The Operating Instructions must be read 1 3 Notes regarding the Operating Instructions Furnace concerned Programat EP 5000 G2 Target group Dental technologists These...

Страница 9: ...structions for Use must be observed The furnace must be operated under the indicated environmental and operating conditions Chapter 9 The Programat EP 5000 G2 must be properly maintained 2 1 1 Risks a...

Страница 10: ...a burn hazard 2 1 12 Contraindication Do not insert any foreign objects into the air vents There is a risk of electrical shock 2 1 13 Contraindication This product contains ceramic fibres and may rele...

Страница 11: ...r must be disconnected to avoid accidental operation Safe operation is no longer possible if the furnace is visibly damaged the furnace does not work the furnace has been stored under unfavourable con...

Страница 12: ...ed by a press drive The firing pressing programs are controlled with the corresponding electronic controls and software Moreover the set and actual temperatures are continuously compared 3 4 Accessori...

Страница 13: ...ents Simply use the two corresponding inserts Fold the side flaps 2 and combine the two packaging parts by means of the transportation flaps The packaging may be disposed with the regular household re...

Страница 14: ...user Lift the furnace head with both hands see picture and carefully position it on the furnace head mounting 43 Secure the cooling tray 34 with the two screws 35 including the silicone washer 47 Plac...

Страница 15: ...aligned with the furnace base mounting mark 42 Make sure that the firing plate holder 48 is not damaged by mounting the furnace head Step 4 Connections Connect the cables of the furnace head with the...

Страница 16: ...blishing additional connections Power connection Please make sure that the voltage indicated on the rating plate complies with the local power supply Connect the power cord with the power socket 17 of...

Страница 17: ...r socket 17 1 Loosen and remove the knurled screw 37 of the hood 36 2 Remove the hood 36 3 Disconnect the thermocouple plug 26 4 Disconnect the heater plug 28 5 Press the leaf spring 32 with a finger...

Страница 18: ...n will be displayed The desired language is set by means of the respective touch button After that the next basic setting screen temperature mode appears These settings are then saved and will no long...

Страница 19: ...Power saving key 89 Power saving function activated only possible with the furnace head closed and the furnace on idle The display shows the power saving icon Pressing any key ends the power saving fu...

Страница 20: ...SP is only active if no active predrying has been selected or if the temperature in the firing chamber is too high when the program is started In addition TSP works only if the stand by temperature of...

Страница 21: ...touch button Laboratory name the keypad is displayed and the name of the laboratory can be entered With Active table and PC the protocol can be sent to the connected PC software instead of being prin...

Страница 22: ...ible with the furnace head closed and the furnace on idle The display shows the power saving icon Pressing any key ends the power saving function The energy saving mode Power Saving Technology is avai...

Страница 23: ...memory stick Select Selection Program manager Copy programs and make a choice using the keys All programs Program copy a single program Group copy a single group When a USB memory stick is operational...

Страница 24: ...e with the ESC key the signal transmitter is also switched off Symbol Name Meaning Symbol Pre vacuum Vacuum generation starts before the heating begins One stage program Abstract firing curve of a one...

Страница 25: ...ettings OSD system 5 8 3 Deactivation In order to deactivate the OSD the brightness has to be set to 0 Colour Activity white blinking The furnace is in the switch on mode self test active white Stand...

Страница 26: ...program has been set to one stage 6 2 2 Operating indicator Firing curve If the program is started with the START key the firing curve display with the vacuum status is shown The following informatio...

Страница 27: ...program group of your choice Now select the desired program e g 1st opaquer Step 2 Now open the furnace head by pressing the Open furnace head key 90 and place the firing tray with the object to be f...

Страница 28: ...etween the Stand by and Parameters screen The shown parameters are displayed for information purposes and cannot be edited Operation screen during a press program in the IPF mode 6 3 4 CDS Crack Detec...

Страница 29: ...TOP key once to pause a running program The green LED in the START key blinks Press the STOP key twice to completely stop the program or press START to continue 6 4 5 Changing the parameters while the...

Страница 30: ...t program function has been activated the furnace head remains open after the firing process the heater is switched off and the green START LED flashes No melody is played if the temperature drops bel...

Страница 31: ...e also observe the shorter service and maintenance intervals What Part When Check all plug in connections for correct fit Var external connections weekly Check if the furnace head opens smoothly and w...

Страница 32: ...Two different chequer board patterns are alternately shown on the entire display This allows the visual check of each individual pix el The display test can be ended by pressing ESC 7 4 Temperature ca...

Страница 33: ...ng movements from the split taper socket 64 with one hand and pull down the press plunger in the firing chamber with the other hand Contraindication Do not touch the thermocouple when replacing the pr...

Страница 34: ...ot to start a firing program 27 no Furnace head cannot be initi alized The furnace head cannot be moved to the final position It might be blocked by an external mechanical source If not contact your S...

Страница 35: ...e memory stick in pro gress memory stick no longer present A program in progress started from the USB memory stick was interrupted by a power failure The program could not be continued since the USB m...

Страница 36: ...head does not open Was the furnace head moved manually Open the furnace head only by using the corresponding keys Switch the furnace on and off again Has the vacuum already been released Is the progr...

Страница 37: ...ions of the closed furnace Depth 470 mm width 305 mm 400 mm with cooling tray Height 565 mm Usable size of the firing chamber Diameter 80 mm Height 48 mm Max firing temperature 1200 C Weight Furnace b...

Страница 38: ...nt Service Center Important information The current program table is also available at www ivoclarvivadent com downloadcenter The program tables can be downloaded from the Internet as PDF files Please...

Страница 39: ...39 10 2 Menu structure 10 2 1 Possibilities of the program selection Firing programs Press programs...

Страница 40: ...40 10 2 3 Overview of program groups 10 2 2 Adjustments Information...

Страница 41: ......

Страница 42: ......

Страница 43: ......

Страница 44: ...3 Piso 14 Col Napoles 03810 M xico D F M xico Tel 52 55 50 62 10 00 Fax 52 55 50 62 10 29 www ivoclarvivadent com mx Ivoclar Vivadent Ltd 12 Omega St Albany PO Box 5243 Wellesley St Auckland New Zeala...

Отзывы: