System-Related Functions
Copyright ©ITECH Electronic Co., Ltd.
61
For example, analog control AC range of 0~350V voltage, when the analog
signal voltage is set to 5V, the voltage ratio set to 50V/1 under source menu, the
instrument output voltage is set to 5*50=250V.
When parallel machine operation, can be controlled through the host analog
interface.
Power Amplification
Analog signals can be input through the analog input interface and power
amplification function can be realized. The operation method is as follows:
Connecting the analog interface, different interface function are different, please
refer to pins description as above.
Voltage and current monitoring
Through the analog interface, the existing output voltage/current can be
monitored. Connect a digital voltmeter or oscilloscope between pin 54
(V_Monitor), pin 6 (I_Monitor) and ground wire 4 (GND) of the analog interface.
The -10V ~ 10V voltage reading corresponds to the power voltage and current
output between
negative full range
and positive full scale
(For AC, 0 to 10V
corresponds to 0 to full scale). The wiring diagram is shown in the figure below.