Inspection and Installation
Copyright ©ITECH Electronic Co., Ltd.
24
⚫
The connection diagram of single phase is shown as follow:
NOTE
When the output voltage has DC voltage, the output terminal L is positive, and N is
negative.
1.
Confirm that the power switch is in the OFF position and verify that there is
no dangerous voltage on the connection terminals.
2.
Remove the output terminals cover of the power system.
3.
Loosen the screws of the output terminals and connect the red and black
test cables to the output terminals. Re-tighten the screws.
When maximum current that one test cable can withstand fails to meet the
rated current, use multiple pieces of red and black test cables. For example,
the maximum current is 1,200A, then 4 pieces of 360A red and black cables
are required.
4.
Thread the red and black test cables through the output terminals cover of
the power system and install the cover.
5.
(Optional) According to the actual situation of DUT, connect the grounding
terminal on the rear panel of the instrument to the DUT to ensure the safe
grounding.
For the location information, see 1.5 Rear Panel Introduction.
6.
Connect the other end of the red and black cables to the DUT. The positive
and negative poles must be properly connected and fastened when wiring.
Connecting the DUT (Remote Sensing)
Remote measurement is available for the following scenarios:
When the DUT consumes large current or the wires are too long, there is a
voltage drop on the wires between DUT and output terminals of the power
system.
To maximize measurement accuracy, the power system provides the remote
measurement terminals VS+ and VS- on the rear panel, which can be used to
measure the terminal voltage of the DUT.
When the power system is used for battery testing in actual applications, the