Technical Specifications
Basic technical data is also visible at the back side sticker of the photometer.
Sensor
silicon photodiode, V
λ
filter
Spectral range
f
0
555 nm -human eye like
Spectral error
< 5 % according to CIE
Full viewing angle
20° (as standard); (5° - 30° ava ilable)
Measuring ranges
0-100; 0-1.000; 0-10.000 cd/m
2
(
0- 65; 0- 650; 0- 6.500 cd/m
2
)
(see calibration)
Accuracy L
20;
linearity
> 3 % +0,5 % of range; > 1 %
Dew & Temperature control
heater 25 W
Sensor-room temperature stability
±1
°C
Temperature range
-40
°C to +50
°C
Power supply
230 V AC ±15 %
Frequency
50 - 60 Hz
Max. control power
2 A / 460 W
Power consumption
< 30 W
Fuse
F 3 A
(5x20 mm)
Casing
IP67 ; Aluminium / Stainless steel / ABS
Size (W x H x D)
395 x 135 (230) x
Ф
145 mm
Weight
2,8 kg
Fixture
wall, X - Y adjustable fixture / pole
-optional
Communication port
RS-232/485 / 9600 bps
Contactor 230 V AC outputs:
Number of channels
4
Output voltage
230 V AC; 0,5 A/CH (16 A peak);
common fuse 3 A
Adjustable range
0 - 10.000 cd/m
2
(0 –6.500 cd/m
2
); auto ranging
Averaging interval
0 - 24 min
5. 1.
Environmental Conditions
Maximum environmental temperature at use is +50
°C; 24 hour average temperature
must not 35
°C. Lowest environmental temperature at use is –40
°C. Relative air
humidity must not exceed 50
% at maximum temperature +50
°C. Higher humidity is
allowed at lower temperature (like: 95
% at +20
°C).
Sea-level elevation must not exceed 2000 metres. When using above 1000 metres
consider lower air dielectricity and different cooling conditions.