Click Clear button to clear all logged data from table and graph. To monitor specific data
closely select the data range in table and click Zoom button.
Data logging is paused by any click on table. To continue set the Auto option in Light
Level Read field of Configuration tab again. Logging is also paused when table is
cleared, zoomed or when data are being exported.
4. Maintenance
4. 1. Checking and Maintenance
It is recommended to:
clean front screen of the photometer (minimum 2 months interval),
clean the housing (1 year interval) and
recalibrate (3 year interval).
Note!
The device requires regular checking and maintenance. Keep your maintenance and
service log up to date. Shorten check intervals if required.
•
Maintenance personnel must be job qualified;
•
Maintenance contractor company must provide education and training;
•
Regular testing and checking must be exercised as described in this manual
and local standards;
•
Any maintenance activity must be noted in the log.
4. 2. Servicing
•
Servicing can only be done by authorised qualified personnel.
•
Keep servicing log.
Servicing for the time of warranty is provided by manufacturer exclusively, after this time it
can be done by approved maintenance contractor.