ISE
Slope Feature & Functions
pH Slope
Press
Slope
key to confirm electrode slope after pH calibration.
The slope displays in the main field and then disappeared.
For the correct operation, the range of slope must be within 80 ~ 120%. If the slope is
not within this range, prefer newly calibrating in order to
prevent the higher error.
It also makes to estimate time of exchange of electrode
since can know error through slope.
ISE Slope
Press
Slope
key to confirm electrode slope after ISE calibration.
The slope displays in the main field and then disappeared.
It also makes to estimate time of exchange of electrode since can know error through
slope.
The slope is displayed with mV unit.
M illivolt / Relative M illivolt M easurement
Meter can measure absolute or relative millivolt. The millivolt modes are useful when
performing potentiometric titration or preparing calibration curves.
Chapter V Calibration and Measurement