20
Temperature Setup
If temperature on display differs from a real temperature, set
a real temperature.
Set temperature by using
or
key.
If finishing setup, press
Setup
or
O ut
key to return to
conductivity initial display.
TDS M ode Setup
In the TDS ready condition if pressing
Setup
key, the display is shown as follows.
TDS factor adjusts by using
or
key and is basically
adjusted to 0.67.
If finishing setup, press
Setup
or
O ut
key to return to TDS
initial display.
Clear data(memory)
pH/ISE/mV/ORP/TEMP Meter If clearing all the stored data, press
M ode
key to
enter mV mode and then press
Select
key to clear.
Therefore, all data which set at setup, is changed to
a basic value.
DO/O2/Air/TEMP Meter If clearing all the stored data, press
M ode
key to
enter O
2
mode and then press
Select
key to clear.
Therefore, all data which set at setup, is changed to
a basic value.
Conductivity/TDS/Salinity/TEMP Meter
If clearing all the stored data, press
M ode
key to
enter Salinity mode and then press
Select
key to
clear. Therefore, all data which set at setup, is
changed to a basic value.
Chapter IV Setup Functions
N eo M et
SETUP
25.0
C
SETUP
0.67
TDS